Journal Information
Vol. 44. Issue. 3.May - June 2024
Pages 313-458
Visits
987
Vol. 44. Issue. 3.May - June 2024
Pages 313-458
Original article
Full text access
Correlation of plasmatic sodium determined by the laboratory and that determined by the dialysis machine
Correlación entre el sodio plasmático determinado por el laboratorio y el determinado por el monitor de hemodiálisis
Visits
987
Joaquim Casalsa, José Jesús Brosetaa,
Corresponding author
jjbroseta@clinic.cat

Corresponding author.
, Rosa María Fernándezb, Diana Rodriguez-Espinosaa, Jimena del Riscoa, Miquel Gómeza, Lida M. Rodasa, Marta Arias-Guilléna, Manel Veraa, Néstor Fontseréa, Naira Ricob, Francisco Maduella
a Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain
b Centro de Diagnóstico Biomédico, Hospital Clínic, Barcelona, Spain
This item has received
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (2)
Abstract
Introduction

Changes in plasma sodium concentration (pNa, expressed in mEq/L) are common in hemodialysis (HD) patients. Hemodialysis monitors can estimate pNa by using an internal algorithm based on ion dialysance measurements. The present study studies the accuracy of the correlation between the pNa estimated by the dialysis monitor and that measured by the biochemistry laboratory at our center.

Material and methods

A single-centre prospective observational study in patients on a chronic HD program with the 6008 CAREsystem monitor and standard sodium (138mmol/L) and bicarbonate (32mmol/L) prescriptions. Venous blood samples were drawn from each patient before and after each HD session to ensure inter- and intra-individual validity. The pNa was measured in the biochemistry laboratory using indirect potentiometry and simultaneously the estimated pNa by the HD monitor was recorded at the beginning and at the end of the HD session. For statistical analysis, a scatterplot was made, and Spearman’s correlation quotient was calculated. In addition, the differences between both methods were represented as Bland-Altman diagrams.

Results

The pre-dialysis pNa measured in the laboratory was 137.49±3.3, and that of the monitor, 137.96±2.91, with a correlation with R2 value of 0.683 (p<0.001). The post-dialysis pNa measured in the laboratory was 137.08±2.23, and that of the monitor was 138.87±1.88, with an R2 of 0.442 (p<0.001). On the Bland-Altman plots, the pre-dialysis pNa has a systematic error of 0.49, in favor of the monitor-estimated pNa, with a 95% confidence interval (CI) of (−3.24 to a 4.22). In the post-dialysis pNa, a systematic error of 1.79 with a 95% CI of (−1.64 to 5.22) was obtained.

Conclusion

The correlation between the pNa estimated by Fresnius 6008 CAREsystem HD monitor and that measured by the laboratory is good, especially pre-dialysis measurements. Further studies should verify the external validity of these results.

Keywords:
Ionic dialysance
Hemodialysis
Individualization
Plasmatic sodium
Sodium module
Resumen
Introducción

Las alteraciones de la concentración plasmática de sodio (Nap, expresado en mEq/L) son frecuentes en pacientes en hemodiálisis (HD). Los monitores de hemodiálisis tienen la capacidad de estimar la Nap mediante un algoritmo interno a partir de las medidas de la dialisancia iónica. En el presente estudio se estuda la correlacionación entre la Nap estimada por el monitor de diálisis y la medida enel laboratorio de bioquímica de nuestro centro.

Material y métodos

Estudio observacional prospectivo y unicéntrico en pacientes en programa crónico de HD con el monitor 6008 CAREsystem y prescripción estándar de sodio (138 mEq/L) y bicarbonato (32mmol/L). De cada paciente se extrajeron muestras de sangre venosa antes y después de la sesión para asegurar validez inter e intraindividual. Se analizó la Nap en el laboratorio mediante potenciometría indirecta y simultáneamente se registraba la estimada por el monitor de HD al inicio y al terminar la sesión. Para el análisis estadístico se realizó un diagrama de dispersión y se calculó el cociente de correlación de Spearman. Además, se representaron las diferencias entre métodos mediante diagramas de Bland-Altman.

Resultados

La Nap prediálisis medida en el laboratorio fue de 137,49±3,3, y el del monitor 137,96±2,91, con una correlación con valor de R2 de 0,683 (p<0,001). La Nap postdiálisis medida en el laboratorio fue de 137,08±2,23 y la del monitor de 138,87±1,88, con una R2 de 0,442 (p<0,001). En los diagramas de Bland-Altman, la Nap prediálisis obtuvo un error sistemático de 0,49 mEq/L a favor de la Nap estimada por el monitor, con un intervalo de confianza (IC) al 95% de (−3,24–4,22). En cuanto a la Nap postdiálisis, se obtuvo un error sistemático de 1,79 mEq/L con un IC al 95% de (−1,64–5,22).

Conclusión

La correlación entre la Nap estimada por el monitor de HD 6008 CAREsystem de Fresenius y la medida por el laboratorio es buena, siendo mejor en las mediciones prediálisis. Nuevos estudios deberán comprobar la validez externa de estos resultados.

Palabras clave:
Dialisancia iónica
Hemodiálisis
individualización
Sodio plasmático
Módulo de sodio
Full Text
Introduction

Sodium (Na) is the most abundant cation in the extracellular compartment and is the most important determinant of plasma volume and osmolarity.1,2 Na homeostasis is essential for maintaining water-electrolyte balance and, ultimately, proper cellular function.3 This depends on the intestinal absorption of Na and its excretion in the feces, sweat and, above all, urine.2,4 For this reason, dysnatremias are frequent in patients with chronic kidney disease and especially in those on hemodialysis (HD).5

Conductivity is considered a reliable and practical surrogate for monitoring the Na concentration of dialysis fluid, so this is used as a safety variable to ensure the mixture of acid concentrate, bicarbonate and treated water is correct.6 A variation of 1mmol/l of Na causes a 0.1mS/cm change in conductivity.

In HD patients, both predialysis hyponatremia (Na < 135mmol/l)7–10 and a difference between final and initial Na of HD treatment (i.e., an Na delta) greater than 4mmol/l11 have been associated with increased mortality. So, it is of great interest to know the plasma sodium concentration (Nap) of HD patients.

In routine clinical practice, monthly or bimonthly determination of Nap, has been performed in the biochemistry laboratory using venous blood samples. However, differences have been observed that should be considered depending on the type of method used in the analysis,12 in particular flame photometry and potentiometry, which includes both direct and indirect potentiometry.

Currently, HD monitors can estimate the patient's Nap, at each session and in real time, by means of an internal algorithm from ionic dialysance measurements. In addition, a recent addition has been introduced in some monitors, an Na module that, from the Nap monitoring of the patient, is able to automatically adjust the Na concentration in the dialysis fluid to try to achieve a neutral diffusive balance, which would approach isonatremia.13–15

Given the disparity of methods for determining Nap in the laboratory and the new possibility of estimating it by ionic dialysance in the HD monitor, the aim of the present study was to evaluate the correlation between Nap estimated by the HD monitor and that determined in the laboratory.

Material and methods

Prospective, single-center, observational study of patients in a chronic HD program. All patients were dialyzed with the 6008 CAREsystem monitor. The acid concentrate prescription and Na (138 mEq/l) and bicarbonate (32mmol/l) concentrations were kept constant in each patient, according to the concentrates' technical data sheet, as well as the rest of the dialytic parameters.

Each patient received between 2 and 5 HD sessions in which pre- and post-dialysis venous blood samples were drawn to ensure inter- and intra-individual validity. Nap was determined in the laboratory using the Atellica™ Solution platform (Siemens), specifically on the LYTE® Integrated Multisensor module that uses indirect potentiometry (IP) in electrolyte measurement. The limit of quantification (LoQ) of Na in this analyzer is less than 50mmol/l (50 mEq/l) with a total error ≤ 20% in serum and plasma. This detection capability was determined in accordance with CLSI document EP17-A2.16

The results are expressed as mean±standard deviation when the variables are quantitative and as absolute and relative frequencies when they are qualitative. To evaluate statistical inference, we constructed scatter diagrams and calculated Spearman's correlation coefficient, and used a Bland-Altman diagram to quantify the mean difference between the 2 methods with a confidence interval that includes 95% of the differences between one technique and the other. Finally, we set the cut-off point for statistical significance at p<0.05. Statistical analysis was performed with the SPSS program® version 25 (IBM Corp., Armonk, NY, USA).

Results

Eighty-one patients (49 males and 32 females) with a mean age of 68±16 years were included. The etiology of renal failure was: chronic glomerulonephritis in 12 cases (14.6%), tubulointerstitial nephropathy in 7 cases (8.5%), vascular etiology in 25 cases (30.5%), polycystic in 6 cases (7.3%), diabetic nephropathy in 15 cases (18.3%), systemic disease in 1 case (1.2%), urological in 3 cases (3.7%), renal tumor in 2 cases (2.4%), cardiorenal in 1 case (1.2%) and cases of non-affiliated etiology in 9 cases (11%). The patients were dialyzed through an arteriovenous fistula in 70% of the cases, through a prosthetic fistula in 5% and the remaining 25% through a tunneled central venous catheter. The different parameters of the dialysis sessions were as follows: dialysis time, 319±72min (240–480min); blood flow (Qb) 411±37min (300–450), dialysis bath flow (Qd) 400ml/min; hemodiafiltration modality in 95% and extended HD in the remaining 5%. Finally, anticoagulation was performed with sodium heparin in 36.5% of cases, low molecular weight heparin in 51.5% of cases and no heparin in the remaining 12%.

A total of 277 dialysis sessions, performed with 15 different 6008 CAREsystem monitors, were analyzed, in which pre- and post-dialysis Nap values, both those determined by the dialysis monitor and those obtained by indirect potentiometry in the laboratory, were collected for comparison.

The predialysis Nap measured in the laboratory was 137.49±3.3 (range: 126–146), while that of the monitor was 137.96±2.91 (range: 128–144). As for post-dialysis, that determined by the laboratory was 137.08±2.23 (131–145) and that of the monitor, 138.87±1.88 (133–144). A correlation was obtained between predialysis Nap measured by both methods, governed by the equation laboratory predialysis Nap=0.938 × Nap monitor predialysis +8.1, with an R2 value of 0.683 and a p<0.001 (Fig. 1A). As for post-dialysis Nap, a correlation was obtained governed by the formula Nap post-laboratory dialysis=0.789 × Nap post-monitor dialysis + 27.6, with an R value2 of 0.442 and a p<0.001 (Fig. 1B). To rule out potential biases, we performed a sub-analysis by dialysis shift of the differences found between monitors and found no statistically significant differences.

Figure 1.

Correlation of plasma sodium concentration measurement in the monitor versus the laboratory. A) In pre-dialysis samples. B) Post-dialysis.

(0.16MB).

Using the Bland-Altman plot, in reference to pre-dialysis Nap, a systematic error of +0.49 mEq/l (95% CI: –3.24 to 4.22) was observed in favor of the monitor with respect to the laboratory. In the case of post-dialysis Nap, a systematic error of +1.79 mEq/l (95% CI: –1.64 to 5.22) in favor of the monitor over the laboratory was observed (Fig. 2). We subsequently performed another post hoc analysis to evaluate whether the differences between the Nap measured by the monitor minus that obtained in the laboratory were consistent as a function of the patient's Nap measured in the laboratory, which we analyzed at both times, finding that the difference is greater in the more hyponatremic patients and that, moreover, this difference gradually decreases until it is reversed in those with higher Na:

ΔNap (laboratory-monitor) predialysis=37.773 – 0.271 × Nap laboratory predialysis and
ΔNap (laboratory-monitor) predialysis=62.087 – 0.44 × Nap laboratory postdialysis, with R2 of 0.23 and 0.328, respectively.

Figure 2.

Bland-Altman plot to quantify the mean difference between the two methods of measuring plasma sodium concentration.

(0.17MB).
Discussion

Intradialytic Na balance is a crucial point of treatment. A positive intradialytic Na gradient favors hemodynamic stability and adequate perfusion of vital organs,9 however, it increases plasma osmolarity, thirst and extracellular volume, thus leading to increased blood pressure, greater left ventricular hypertrophy and favoring the development of cardiovascular events.17 In contrast, a negative gradient is associated with lower interdialytic weight gain and better blood pressure control at the expense of favoring intradialytic hypotension and tissue hypoperfusion.9 This is relevant when prescribing HD treatment, since different studies have been published, with consistent results among them, that correlate the Na prescribed on the monitor with the Na finally achieved in the dialysis fluid.18

This observational study has shown that there are no major differences, in our center, between Na measured by the HD machine by ionic dialysance and Na measured by indirect potentiometry in the laboratory from venous blood samples. As shown in Fig. 1, there is a good correlation between the Nap measured by the machine and that measured in the laboratory, with the best fit in the case of predialysis Nap. In general terms, what we observed is that the value of Nap is overestimated by the dialysis machine, more in the high range of Na values and, above all, in the case of post-dialysis Nap.

The Bland-Altman diagram in Fig. 2 shows how the predialysis Nap has a systematic error of 0.49 mEq/l in favor of the monitor. That is, the monitor will give us a slightly higher Nap than the laboratory Na. Assuming 0.5 mEq/l mean systematic error seems a good option, since the laboratory gives us the Nap figures in whole numbers, without decimals, so we would be dealing with a negligible error. However, in the case of post-dialysis Nap, the mean systematic error is 1.79 mEq/l in favor of the monitor. The fact that postdialysis Nap in the different analyses correlates worse than predialysis Nap could be related to being an indirect measurement method based on ionic dialysance. Thus, there could be greater variation in the ionic composition of the plasma at the end of dialysis (other elements are added, such as chlorine or bicarbonate) which could alter the estimation of the Na valuep,13 in addition to other factors such as changes in glycemia,19 protein binding or the formation of complexes with anions such as sulfate and phosphate.20 This means that the post-dialysis Nap value should not be considered reliable in clinical situations that require an exact natremia value.

Both standard errors have a wide 95% confidence interval, which could have clinical implications, although this is lower in the case of post-dialysis Nap. In fact, if we observe the range of Nap values, predialysis, it falls between 127 and 144 mEq/l, whereas post-dialysis it narrows between 134 and 144 mEq/l, a finding that supports the effect of dialysis fluid in modifying Nap values during HD treatment.

Subsequent post hoc analysis confirmed that pre-dialysis and post-dialysis Nap remained without statistically significant differences between the different monitors and dialysis shifts. However, we did find that the differences between the Nap measured by the monitor and that of the laboratory, both pre- and post-dialysis, depended on the patient's blood Nap, with patients with a lower Nap having a greater difference in favor of the monitor, while those with a higher Nap had a smaller difference and in favor of the laboratory. These findings, which become clinically relevant in extreme Nap, should be considered. Perhaps this margin of error could be corrected by the formula suggested in this work, especially in hyponatremic patients, to correct the monitor's overestimate.

A factor to consider when externally validating this study is the method used for Na measurement in each laboratory. Currently the main methods used for Nap measurement or serum are flame atomic emission spectrometry and ion selective electrode potentiometry, the former being the reference method so far. In this method, the sample in solution is nebulized and introduced into the flame, where it is atomized. As the excited atoms decay to the basal electronic state, radiation is emitted which passes through a monochromator that isolates the characteristic wavelength for the desired ion. The intensity of light emitted will be proportional to the concentration of the ion present in the solution. Although the reference method for measuring Nap or serum has always been atomic emission spectrometry, this is a laborious method that does not allow rapid sample throughput. For this reason, most laboratories use potentiometry as the main method, since the technology used is much simpler and allows automation of the samples. It consists of measuring the electrical potential between a reference electrode and an indicator electrode in an electrochemical cell when no current is flowing. The reference electrode is immersed in a solution of known concentration that generates a constant potential, while the indicator electrode, immersed in the test solution, is a membrane electrode selective to the ion to be determined. The potential generated at this electrode will vary according to the concentration of the ion present in the solution.21 The potentiometer of the electrochemical cell is then responsible for determining the potential difference between the two electrodes and then, by applying the Nernst equation,22 it is possible to calculate the concentration of the ion in question.

This method in turn is classified into two types, direct potentiometry (DP) and indirect potentiometry (IP). In the first case, it is not necessary to dilute the sample, and it is the method used in gas analyzers located, mainly, in the place of patient care.23 In the case of IP, as in flame photometry, the sample is diluted, which allows the measured ion activity to be closer to the ion concentration and, for this reason, its use is more standardized than that of PD.16

The PI and PD methods measure the electrolyte activities in the aqueous phase of the plasma, which represents about 93%, but assume this result as if it were the concentration of electrolyte present in the total plasma, assuming a normal solid phase of about 7% which is basically composed of proteins and lipids. They differ only in that in the IP the percentage of solid phase component is considered for the calculation of the Na concentration, as long as the water content in the aqueous phase of the plasma remains constant, this difference between the Na ion concentration in the total plasma and the Na ion concentration in the aqueous phase of the plasma is predictable and can be ignored. However, the problem arises in some clinical conditions, such as hyperlipidemia or hyperproteinemia, in which the water content is replaced by protein or lipid. In these situations, the aqueous phase of plasma decreases and the solid phase increases and causes a phenomenon known as pseudohyponatremia.24 This is even more pronounced with sample dilution, because a smaller volume of plasma may be aspirated than expected. This phenomenon, however, is not expected in PD because it neglects the solid phase component. For this reason, the European Society of Endocrinology (ESE), the European Society of Intensive Care Medicine (ESICM) and the European Renal Association (ERA) recommend using PD test results to diagnose dysnatremias.

In subsequent studies it would be important to determine the Na concentration by PD in those HD patients with a high concentration of proteins and/or lipids to assess whether these differences are still observed with the Nap measured by ionic dialysance or whether, on the contrary, the discrepancies are smaller. Furthermore, it is important to point out that dialysis monitors of different brands, with identical conductivity prescription, obtain different Na concentrations, with a range exceeding 4mmol/l. This is due to the temperature coefficient chosen to correct conductivity to the ISO standard of 25 C°.25 Another limitation of this work is that we did not correct natremia for glycemia.

In any case, our results agree with those obtained by Maierhofer et al. who, in a study including 384 dialysis sessions performed in 75 different patients, demonstrated a good correlation between pre-dialysis Nap concentration determined by the monitor compared to blood samples analyzed by direct potentiometry.26

In conclusion, the present study is a proof of concept indicating a good correlation between Nap measured by the Fresenius CAREsystem 6008 HD monitor versus the laboratory. However, this is a proof of concept, and caution should be exercised in its interpretation, especially in cases where the result is above normal values and in post-dialysis samples. Further studies evaluating Nap measured by the monitor against the diagnostic gold standard are needed.

Financing

This study has not received financial support.

Conflict of interest

FM has received fees from Amgen, Baxter, Fresenius Medical Care, Medtronic, Nipro and Vifor. JJB has received honoraria from Fresenius Medical Care. The rest of the authors declare no conflicts of interest.

Acknowledgments

We would like to express our gratitude to all the patients who participated, as well as to all the staff of the Dialysis Section of the Hospital Clínic de Barcelona for their collaboration and enthusiasm in this study.

References
[1]
P. Strazzullo, C. Leclercq.
Sodium.
Adv Nutr, 5 (2014), pp. 188-190
[2]
P. Bie.
Mechanisms of sodium balance: Total body sodium, surrogate variables, and renal sodium excretion.
Am J Physiol Regul Integr Comp Physiol, 315 (2018), pp. R945-R962
[3]
A. Bernal, M.A. Zafra, M.J. Simón, J. Mahía.
Sodium Homeostasis, a Balance Necessary for Life.
Nutrients, 15 (2023), pp. 395
[4]
M. Noda, T. Matsuda.
Central regulation of body fluid homeostasis.
Proc Jpn Acad Ser B Phys Biol Sci, 98 (2022), pp. 283-324
[5]
H.A. Koomans, J.C. Roos, E.J. Dorhout Mees, I.M.K. Delawi.
Sodium balance in renal failure. A comparison of patients with normal subjects under extremes of sodium intake.
Hypertension, 7 (1985), pp. 714-721
[6]
F. Maduell, J.J. Broseta, J. Casals, M. Gómez, L.M. Rodas, M. Arias-Guillén, et al.
Variaciones de la conductividad y cambios en el sodio plasmático durante hemodiálisis relacionado con el cambio de monitor.
[7]
N. Sato, M. Gheorghiade, K. Kajimoto, R. Munakata, Y. Minami, M. Mizuno, et al.
Hyponatremia and in-hospital mortality in patients admitted for heart failure (from the ATTEND registry).
Am J Cardiol, 111 (2013), pp. 1019-1025
[8]
S.S. Waikar, D.B. Mount, G.C. Curhan.
Mortality after hospitalization with mild, moderate, and severe hyponatremia.
Am J Med, 122 (2009), pp. 857-865
[9]
R. Pérez-García, I. Palomares, J.I. Merello, R. Ramos, F. Maduell, M. Molina, et al.
Hyponatraemia, mortality and haemodialysis: An unexplained association.
Nefrologia, 36 (2016), pp. 42-50
[10]
X. Ye, J.P. Kooman, F.M. Van Der Sande, B. Canaud, S. Stuard, M. Etter, et al.
Increased Mortality Associated with Higher Pre-Dialysis Serum Sodium Variability: Results of the International MONitoring Dialysis Outcome Initiative.
Am J Nephrol, 49 (2019), pp. 1-10
[11]
K. Fujisaki, N. Joki, S. Tanaka, E. Kanda, T. Hamano, I. Masakane, et al.
Pre-dialysis Hyponatremia and Change in Serum Sodium Concentration During a Dialysis Session Are Significant Predictors of Mortality in Patients Undergoing Hemodialysis.
Kidney Int Rep, 6 (2020), pp. 342-350
[12]
R.A. Garcia, C.P. Vanelli, O. dos Santos Pereira Junior, J.O. do Amaral Corrêa.
Comparative analysis for strength serum sodium and potassium in three different methods: Flame photometry, ion-selective electrode (ISE) and colorimetric enzymatic.
J Clin Lab Anal, 32 (2018),
[13]
U. Kuhlmann, A. Maierhofer, B. Canaud, J. Hoyer, M. Gross.
Zero Diffusive Sodium Balance in Hemodialysis Provided by an Algorithm-Based Electrolyte Balancing Controller: A Proof of Principle Clinical Study.
Artif Organs, 43 (2019), pp. 150-158
[14]
M. Ságová, R. Wojke, A. Maierhofer, M. Gross, B. Canaud, A. Gauly.
Automated individualization of dialysate sodium concentration reduces intradialytic plasma sodium changes in hemodialysis.
Artif Organs, 43 (2019), pp. 1002-1013
[15]
F. Maduell, J.J. Broseta, D. Rodríguez-Espinosa, J. Casals, V. Escudero, M. Gomez, et al.
Practical implementation and clinical benefits of the new automated dialysate sodium control biosensor.
Clin Kidney J, 23 (2023),
[16]
S. Malandrini, S.A.G. Lava, M.G. Bianchetti, F. Meani, P.B. Faré, P. Camozzi, et al.
Which laboratory technique is used for the blood sodium analysis in clinical research? A systematic review.
[17]
N.R. Cook, J.A. Cutler, E. Obarzanek, J.E. Buring, K.M. Rexrode, S.K. Kumanyika, et al.
Long term effects of dietary sodium reduction on cardiovascular disease outcomes: observational follow-up of the trials of hypertension prevention (TOHP).
[18]
J.K.C. Ng, B. Smyth, M.R. Marshall, A. Kang, J. Pinter, A. Bassi, et al.
Relationship between measured and prescribed dialysate sodium in haemodialysis: A systematic review and meta-analysis.
Nephrol Dial Transplant, 36 (2021), pp. 695-703
[19]
M. Albalate Ramón, M. Albalate, P. de Sequera Ortiz, M.J. Ruiz Alvarez, E. Corchete, T. Talavan, et al.
Set-point de sodio en hemodiálisis: ¿es lo que vemos en la clínica?.
Nefrología, 33 (2013), pp. 808-815
[20]
J.G. Raimann, S. Thijssen, L.A. Usvyat, N.W. Levin, P. Kotanko.
Sodium Alignment in Clinical Practice-Implementation and Implications.
[21]
N.F. MacDonald, P.Z. Williams, J.I. Burton, J.G. Batsakis.
Sodium and potassium measurements. Direct potentiometry and flame photometry.
Am J Clin Pathol, 76 (1981),
[22]
11.2: Potentiometric Methods - Chemistry LibreTexts [accessed May 1 2023] Available from: https://chem.libretexts.org/Courses/Northeastern_University/11%3A_Electrochemical_Methods/11.2%3A_Potentiometric_Methods.
[23]
S.A.G. Lava, M.G. Bianchetti, G.P. Milani.
Testing Na+ in blood.
Clin Kidney J, 10 (2017), pp. 147-148
[24]
G.H. Kim.
Pseudohyponatremia: does it matter in current clinical practice?.
Electrolyte Blood Press, 4 (2006), pp. 77-82
[25]
A. Stragier, F. Lopot, F. Švára, V. Polakovič.
Fallacies and Pitfalls of Dialysis Sodium Prescription and Control.
Blood Purif, 46 (2018), pp. 27-33
[26]
A. Maierhofer, B. Pinto, R. Wojke, P. Ponce.
FP546 Conductivity based online estimation of predialytic plasma sodium: clinical assessment.
Nephrol Dial Transplant, 34 (2019),
Copyright © 2023. Sociedad Española de Nefrología
Download PDF
Idiomas
Nefrología (English Edition)
Article options
Tools
es en

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?