array:22 [
  "pii" => "X0211699514054894"
  "issn" => "02116995"
  "doi" => "10.3265/Nefrologia.pre2014.Jul.12682"
  "estado" => "S300"
  "fechaPublicacion" => "2014-11-01"
  "documento" => "article"
  "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
  "subdocumento" => "fla"
  "cita" => "Nefrologia. 2014;34:698-702"
  "abierto" => array:3 [
    "ES" => true
    "ES2" => true
    "LATM" => true
  ]
  "gratuito" => true
  "lecturas" => array:2 [
    "total" => 17966
    "formatos" => array:3 [
      "EPUB" => 373
      "HTML" => 16178
      "PDF" => 1415
    ]
  ]
  "Traduccion" => array:1 [
    "en" => array:17 [
      "pii" => "X2013251414054891"
      "issn" => "20132514"
      "doi" => "10.3265/Nefrologia.pre2014.Jul.12682"
      "estado" => "S300"
      "fechaPublicacion" => "2014-11-01"
      "documento" => "article"
      "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
      "subdocumento" => "fla"
      "cita" => "Nefrologia (English Version). 2014;34:698-702"
      "abierto" => array:3 [
        "ES" => true
        "ES2" => true
        "LATM" => true
      ]
      "gratuito" => true
      "lecturas" => array:2 [
        "total" => 24236
        "formatos" => array:3 [
          "EPUB" => 428
          "HTML" => 22113
          "PDF" => 1695
        ]
      ]
      "en" => array:10 [
        "idiomaDefecto" => true
        "titulo" => "Hypersensitivity reactions to synthetic haemodialysis membranes ' an emerging issue?"
        "tienePdf" => "en"
        "tieneTextoCompleto" => "en"
        "tieneResumen" => "en"
        "paginas" => array:1 [
          0 => array:2 [
            "paginaInicial" => "698"
            "paginaFinal" => "702"
          ]
        ]
        "contieneResumen" => array:1 [
          "en" => true
        ]
        "contieneTextoCompleto" => array:1 [
          "en" => true
        ]
        "contienePdf" => array:1 [
          "en" => true
        ]
        "autores" => array:1 [
          0 => array:2 [
            "autoresLista" => "M. Antonia Álvarez-de Lara, Alejandro Martín-Malo"
            "autores" => array:2 [
              0 => array:2 [
                "nombre" => "M. Antonia"
                "apellidos" => "Álvarez-de Lara"
              ]
              1 => array:2 [
                "nombre" => "Alejandro"
                "apellidos" => "Martín-Malo"
              ]
            ]
          ]
        ]
      ]
      "idiomaDefecto" => "en"
      "Traduccion" => array:1 [
        "es" => array:9 [
          "pii" => "X0211699514054894"
          "doi" => "10.3265/Nefrologia.pre2014.Jul.12682"
          "estado" => "S300"
          "subdocumento" => ""
          "abierto" => array:3 [
            "ES" => true
            "ES2" => true
            "LATM" => true
          ]
          "gratuito" => true
          "lecturas" => array:1 [
            "total" => 0
          ]
          "idiomaDefecto" => "es"
          "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699514054894?idApp=UINPBA000064"
        ]
      ]
      "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X2013251414054891?idApp=UINPBA000064"
      "url" => "/20132514/0000003400000006/v0_201502091558/X2013251414054891/v0_201502091558/en/main.assets"
    ]
  ]
  "itemSiguiente" => array:17 [
    "pii" => "X0211699514054886"
    "issn" => "02116995"
    "doi" => "10.3265/Nefrologia.pre2014.Jul.12548"
    "estado" => "S300"
    "fechaPublicacion" => "2014-11-01"
    "documento" => "article"
    "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
    "subdocumento" => "fla"
    "cita" => "Nefrologia. 2014;34:703-9"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 11279
      "formatos" => array:3 [
        "EPUB" => 351
        "HTML" => 9875
        "PDF" => 1053
      ]
    ]
    "es" => array:10 [
      "idiomaDefecto" => true
      "titulo" => "Disfunción sexual en pacientes en diálisis peritoneal: relación con factores metabólicos y con la percepción de la calidad de vida"
      "tienePdf" => "es"
      "tieneTextoCompleto" => 0
      "tieneResumen" => array:2 [
        0 => "es"
        1 => "en"
      ]
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "703"
          "paginaFinal" => "709"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Sexual dysfunction in men and women on peritoneal dialysis: Differential link with metabolic factors and quality of life perception"
        ]
      ]
      "contieneResumen" => array:2 [
        "es" => true
        "en" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "Pedro Azevedo, Ricardo Santos, José Durães, Olívia Santos, Maria J. Carvalho, António Cabrita, Anabela Rodrigues"
          "autores" => array:7 [
            0 => array:2 [
              "nombre" => "Pedro"
              "apellidos" => "Azevedo"
            ]
            1 => array:2 [
              "nombre" => "Ricardo"
              "apellidos" => "Santos"
            ]
            2 => array:2 [
              "nombre" => "José"
              "apellidos" => "Durães"
            ]
            3 => array:2 [
              "nombre" => "Olívia"
              "apellidos" => "Santos"
            ]
            4 => array:2 [
              "nombre" => "Maria"
              "apellidos" => "J. Carvalho"
            ]
            5 => array:2 [
              "nombre" => "António"
              "apellidos" => "Cabrita"
            ]
            6 => array:2 [
              "nombre" => "Anabela"
              "apellidos" => "Rodrigues"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "Traduccion" => array:1 [
      "en" => array:9 [
        "pii" => "X2013251414054883"
        "doi" => "10.3265/Nefrologia.pre2014.Jul.12548"
        "estado" => "S300"
        "subdocumento" => ""
        "abierto" => array:3 [
          "ES" => true
          "ES2" => true
          "LATM" => true
        ]
        "gratuito" => true
        "lecturas" => array:1 [
          "total" => 0
        ]
        "idiomaDefecto" => "en"
        "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X2013251414054883?idApp=UINPBA000064"
      ]
    ]
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699514054886?idApp=UINPBA000064"
    "url" => "/02116995/0000003400000006/v0_201502091330/X0211699514054886/v0_201502091331/es/main.assets"
  ]
  "itemAnterior" => array:17 [
    "pii" => "X0211699514054909"
    "issn" => "02116995"
    "doi" => "10.3265/Nefrologia.pre2014.Jul.12610"
    "estado" => "S300"
    "fechaPublicacion" => "2014-11-01"
    "documento" => "article"
    "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
    "subdocumento" => "fla"
    "cita" => "Nefrologia. 2014;34:693-7"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 53290
      "formatos" => array:3 [
        "EPUB" => 347
        "HTML" => 51440
        "PDF" => 1503
      ]
    ]
    "es" => array:10 [
      "idiomaDefecto" => true
      "titulo" => "Doctor, ¿cuánto debo beber?"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "693"
          "paginaFinal" => "697"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Doctor, how much should I drink?"
        ]
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "resumenGrafico" => array:2 [
        "original" => 0
        "multimedia" => array:8 [
          "identificador" => "fig1"
          "etiqueta" => "Tab.  1"
          "tipo" => "MULTIMEDIAFIGURA"
          "mostrarFloat" => true
          "mostrarDisplay" => false
          "copyright" => "Elsevier España"
          "figura" => array:1 [
            0 => array:4 [
              "imagen" => "12610_19904_59018_es_12610_t1.jpg"
              "Alto" => 98
              "Ancho" => 875
              "Tamanyo" => 86461
            ]
          ]
          "descripcion" => array:1 [
            "es" => "Parámetros urinarios en una serie de 94 pacientes con enfermedad renal crónica avanzada estadios 4 y 5 de nuestra consulta"
          ]
        ]
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "Víctor Lorenzo"
          "autores" => array:1 [
            0 => array:2 [
              "nombre" => "Víctor"
              "apellidos" => "Lorenzo"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "Traduccion" => array:1 [
      "en" => array:9 [
        "pii" => "X2013251414054906"
        "doi" => "10.3265/Nefrologia.pre2014.Jul.12610"
        "estado" => "S300"
        "subdocumento" => ""
        "abierto" => array:3 [
          "ES" => true
          "ES2" => true
          "LATM" => true
        ]
        "gratuito" => true
        "lecturas" => array:1 [
          "total" => 0
        ]
        "idiomaDefecto" => "en"
        "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X2013251414054906?idApp=UINPBA000064"
      ]
    ]
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699514054909?idApp=UINPBA000064"
    "url" => "/02116995/0000003400000006/v0_201502091330/X0211699514054909/v0_201502091331/es/main.assets"
  ]
  "asociados" => array:1 [
    0 => array:15 [
      "pii" => "XX342164212000352"
      "issn" => "X3421642"
      "doi" => "10.3265/Nefrologia.2010.pub1.ed80.chapter2810"
      "estado" => "S300"
      "fechaPublicacion" => "2012-12-01"
      "documento" => "article"
      "subdocumento" => "fla"
      "cita" => "Nefrología al día. 2012;7:0"
      "abierto" => array:3 [
        "ES" => true
        "ES2" => true
        "LATM" => true
      ]
      "gratuito" => true
      "lecturas" => array:2 [
        "total" => 65919
        "formatos" => array:3 [
          "EPUB" => 559
          "HTML" => 59910
          "PDF" => 5450
        ]
      ]
      "es" => array:11 [
        "idiomaDefecto" => true
        "titulo" => "Dializadores y membranas de diálisis"
        "tienePdf" => "es"
        "tieneTextoCompleto" => "es"
        "tieneResumen" => "es"
        "paginas" => array:1 [
          0 => array:1 [
            "paginaInicial" => "0"
          ]
        ]
        "contieneResumen" => array:1 [
          "es" => true
        ]
        "contieneTextoCompleto" => array:1 [
          "es" => true
        ]
        "contienePdf" => array:1 [
          "es" => true
        ]
        "resumenGrafico" => array:2 [
          "original" => 0
          "multimedia" => array:8 [
            "identificador" => "fig1"
            "etiqueta" => "Figura 1"
            "tipo" => "MULTIMEDIAFIGURA"
            "mostrarFloat" => true
            "mostrarDisplay" => false
            "copyright" => "Elsevier España"
            "figura" => array:1 [
              0 => array:4 [
                "imagen" => "1847_120_336_es_t1_dmd.jpg"
                "Alto" => 696
                "Ancho" => 900
                "Tamanyo" => 178459
              ]
            ]
            "descripcion" => array:1 [
              "es" => "Membranas de diàlisis"
            ]
          ]
        ]
        "autores" => array:1 [
          0 => array:2 [
            "autoresLista" => "Alejandro Martín Malo, ALM de Francisco"
            "autores" => array:2 [
              0 => array:2 [
                "nombre" => "Alejandro"
                "apellidos" => "Martín Malo"
              ]
              1 => array:2 [
                "nombre" => "ALM"
                "apellidos" => "de Francisco"
              ]
            ]
          ]
        ]
      ]
      "idiomaDefecto" => "es"
      "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/XX342164212000352?idApp=UINPBA000064"
      "url" => "/X3421642/0000000700000001/v0_201502191227/XX342164212000352/v0_201502191229/es/main.assets"
    ]
  ]
  "es" => array:10 [
    "idiomaDefecto" => true
    "titulo" => "Hipersensibilidad a membranas sintéticas de hemodiálisis, ¿un problema emergente?"
    "tieneTextoCompleto" => true
    "paginas" => array:1 [
      0 => array:2 [
        "paginaInicial" => "698"
        "paginaFinal" => "702"
      ]
    ]
    "autores" => array:1 [
      0 => array:3 [
        "autoresLista" => "M. Antonia Álvarez-de Lara, Alejandro Martín-Malo"
        "autores" => array:2 [
          0 => array:4 [
            "nombre" => "M. Antonia"
            "apellidos" => "Álvarez-de Lara"
            "email" => array:1 [
              0 => "malvarezlaras@senefro.org"
            ]
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "affa"
              ]
            ]
          ]
          1 => array:3 [
            "nombre" => "Alejandro"
            "apellidos" => "Mart&#237;n-Malo"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "affa"
              ]
            ]
          ]
        ]
        "afiliaciones" => array:1 [
          0 => array:3 [
            "entidad" => "Servicio de Nefrología, Hospital Universitario Reina Sofía, Córdoba,   "
            "etiqueta" => "<span class="elsevierStyleSup">a</span>"
            "identificador" => "affa"
          ]
        ]
      ]
    ]
    "titulosAlternativos" => array:1 [
      "en" => array:1 [
        "titulo" => "Hypersensitivity reactions to synthetic haemodialysis membranes &#191; an emerging issue&#63;"
      ]
    ]
    "textoCompleto" => "<p class="elsevierStylePara"><span class="elsevierStyleBold">INTRODUCCI&#211;N</span></p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara">Las reacciones de tipo anafil&#225;ctico o anafilactoide que aparecen durante la hemodi&#225;lisis &#40;HD&#41; se conocen desde 1975<span class="elsevierStyleSup">1</span>&#46; Desde entonces se han publicado numerosos art&#237;culos en los que se comunican este tipo de reacciones&#44; aunque la mayor&#237;a de ellos con pocos casos y con diferentes membranas&#44; agentes esterilizantes&#44; medicaci&#243;n administrada durante la di&#225;lisis&#44; grado de pureza del agua&#44; etc&#46; No hay estudios epidemiol&#243;gicos prospectivos bien dise&#241;ados que nos permitan conocer la incidencia exacta y el alcance de este tipo de reacciones&#46; En un estudio realizado en 1985&#44; Daugirdas comunic&#243; 21 reacciones severas&#44; con un caso fatal&#44; en 260&#160;000 sesiones de HD<span class="elsevierStyleSup">2</span>&#46; En 1987 se investig&#243; la prevalencia de reacciones de hipersensibilidad &#171;de primer uso&#187; en Reino Unido y se encontr&#243; que 1 de cada 20 a 50 pacientes era susceptible de presentar una reacci&#243;n anafilactoide con un nuevo dializador&#44; lo cual indica la magnitud del problema<span class="elsevierStyleSup">3</span>&#46; Estas reacciones no se relacionaban con una membrana o tipo de dializador o con una t&#233;cnica de di&#225;lisis determinados&#46; A&#241;os despu&#233;s&#44; otro estudio analiz&#243; la incidencia de reacciones en 1536 pacientes de 30 centros de di&#225;lisis &#40;122&#160;694 sesiones&#41; y se observ&#243; una incidencia anual de 0&#44;17 por 1000 sesiones con membranas celul&#243;sicas y 4&#44;2 por 1000 sesiones con membranas sint&#233;ticas<span class="elsevierStyleSup">4</span>&#46; Esto indica que&#44; aunque no son extremadamente frecuentes&#44; las reacciones aparecen de vez en cuando en todas las unidades de di&#225;lisis y que&#44; aparentemente&#44; es m&#225;s frecuente que est&#233;n asociadas al uso de membranas sint&#233;ticas&#46;</p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">TIPOS DE REACCI&#211;N</span></p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara">Las reacciones que aparecen durante la HD son el resultado de una respuesta inmunoal&#233;rgica por parte del paciente tras la exposici&#243;n a sustancias extra&#241;as al organismo que est&#225;n presentes en el circuito extracorp&#243;reo y&#47;o por la respuesta inducida por la interacci&#243;n de la sangre con la membrana del dializador<span class="elsevierStyleSup">5</span>&#46; Pueden ser de dos tipos<span class="elsevierStyleSup">6</span>&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">&#160;</span></p><p class="elsevierStylePara"><span class="elsevierStyleBold">Tipo A o reacciones de hipersensibilidad</span></p><p class="elsevierStylePara"><span class="elsevierStyleItalic">&#160;</span></p><p class="elsevierStylePara">Se presentan normalmente en los primeros minutos de la di&#225;lisis&#44; aunque pueden aparecer hasta 30 minutos despu&#233;s del inicio&#46; Se caracterizan por urticaria&#44; tos&#44; rinorrea&#44; lagrimeo&#44; calambres abdominales&#44; prurito&#44; sensaci&#243;n de quemaz&#243;n&#44; angioedema&#44; disnea e incluso colapso circulatorio y muerte&#46; Son reacciones graves que obligan a parar inmediatamente la di&#225;lisis&#44; siendo aconsejable no retornar la sangre del circuito extracorp&#243;reo&#46; Estas reacciones de tipo A pueden ser&#44; a su vez&#44; anafil&#225;cticas cuando est&#225;n mediadas por la IgE o anafilactoides si no est&#225;n mediadas por la IgE&#46;</p><p class="elsevierStylePara">Las m&#225;s caracter&#237;sticas se han descrito en relaci&#243;n con el &#243;xido de etileno &#40;OE&#41;&#44; el re&#250;so de dializadores y por la combinaci&#243;n de membranas de poliacrilonitrilo &#40;PAN&#44; AN69&#41; e inhibidores de la enzima conversora de la angiotensina &#40;IECA&#41;&#46;</p><p class="elsevierStylePara">La cl&#225;sica reacci&#243;n al OE se daba exclusivamente durante el primer uso de un dializador esterilizado con OE que no hab&#237;a sido cebado adecuadamente&#46; En la mayor&#237;a de los casos exist&#237;a una IgE frente al OE detectable en el plasma<span class="elsevierStyleBold"> </span>de los pacientes que sufr&#237;an este tipo de reacciones<span class="elsevierStyleSup">1&#44;7</span>&#46; En la actualidad este agente esterilizante ya no se utiliza en los dializadores y ha sido sustituido por radiaci&#243;n gamma o vapor&#46; De todos modos&#44; ante una reacci&#243;n en di&#225;lisis hay que tener presente que el OE a&#250;n se utiliza para esterilizar algunas agujas&#44; jeringas y l&#237;neas de di&#225;lisis&#46; Tambi&#233;n se han descrito reacciones anafilactoides cuando se reutilizan dializadores de polisulfona y acetato de celulosa<span class="elsevierStyleSup">8</span>&#44; as&#237; como un aumento del riesgo de aparici&#243;n de reacciones de hipersensibilidad cuando se usaba hipoclorito s&#243;dico o per&#243;xido de hidr&#243;geno para lavar el compartimento sangu&#237;neo de los dializadores<span class="elsevierStyleSup">9</span>&#46;</p><p class="elsevierStylePara">En los a&#241;os noventa se describi&#243; la aparici&#243;n de reacciones anafilactoides en pacientes dializados con membrana de AN69 &#40;PAN&#41; y que recib&#237;an&#44; al mismo tiempo&#44; tratamiento con IECA<span class="elsevierStyleSup">4&#44;8&#44;10&#44;11</span>&#46; Eran reacciones de tipo A&#44; por el momento de su aparici&#243;n &#40;en los primeros minutos de la di&#225;lisis&#41; y por la sintomatolog&#237;a de los pacientes&#46; El mediador de estas reacciones es la bradikinina&#46; La superficie de la membrana de AN69 tiene una carga el&#233;ctrica fuertemente negativa capaz de activar el sistema de contacto e inducir la producci&#243;n de factor de Hageman&#46; Este convierte la prekalicre&#237;na en kalicre&#237;na&#44; que act&#250;a sobre el kinin&#243;geno para liberar bradikinina&#46; Los IECA&#44; que inhiben la degradaci&#243;n de la bradikinina&#44; hacen que esta se acumule en sangre hasta un nivel de 20 a 30 veces superior al normal&#44; facilitando la aparici&#243;n de los s&#237;ntomas de anafilaxia<span class="elsevierStyleSup">12&#44;13</span>&#46; El recubrimiento de la superficie de la membrana de AN69 con un pol&#237;mero biocompatible &#40;SPAN&#44; AN69 ST&#41; proporciona una neutralizaci&#243;n parcial de la electronegatividad y una reducci&#243;n en la producci&#243;n de bradikinina&#46; Con esta modificaci&#243;n en la membrana se ha podido dializar con AN69 ST a pacientes tratados con IECA y con historia de reacci&#243;n anafilactoide durante la HD con AN69&#44; sin que presentasen ning&#250;n problema<span class="elsevierStyleSup">14</span>&#46;</p><p class="elsevierStylePara">Por &#250;ltimo&#44; tambi&#233;n pueden causar reacciones de hipersensibilidad otros compuestos utilizados en las unidades de di&#225;lisis como el formaldeh&#237;do<span class="elsevierStyleSup">15</span>&#44; el l&#225;tex<span class="elsevierStyleSup">16</span>&#44; la heparina<span class="elsevierStyleSup">17&#44;18</span> y el hierro intravenoso<span class="elsevierStyleSup">19</span>&#46; Esto hace que&#44; con frecuencia&#44; ante una reacci&#243;n de tipo A durante la HD sea dif&#237;cil encontrar el agente causal&#46;</p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">Tipo B o inespec&#237;ficas</span></p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara">Son m&#225;s frecuentes y menos graves que las de tipo A&#46; Los s&#237;ntomas son dolor tor&#225;cico&#44; disnea&#44; n&#225;useas&#44; v&#243;mitos e hipotensi&#243;n&#46; Son m&#225;s tard&#237;as&#44; aparecen a los 15-30 minutos de iniciada la di&#225;lisis&#44; aunque pueden hacerlo incluso m&#225;s tarde&#44; y generalmente se resuelven a medida que transcurre la sesi&#243;n&#44; sin necesidad de desconectar al paciente&#46; Son debidas a una leucostasis pulmonar secundaria a la activaci&#243;n del complemento por la membrana de di&#225;lisis que genera C3a y C5a&#46; Los grupos hidroxilo libres de la membrana de di&#225;lisis activan la v&#237;a alterna del complemento&#44; gener&#225;ndose las anafilotoxinas C3a y C5a&#46; Esta &#250;ltima se une a los receptores de membrana de los leucocitos produciendo activaci&#243;n&#44; agregaci&#243;n y adherencia de estos al endotelio del capilar pulmonar&#44; lo que da lugar a las ya mencionadas leucostasis pulmonar&#44; leucopenia e hipoxemia<span class="elsevierStyleSup">20-23</span>&#46; El nadir de la leucopenia se produce a los 15 minutos de iniciada la sesi&#243;n y posteriormente la cifra de leucocitos en sangre va aumentando y recupera el valor predi&#225;lisis a la hora aproximadamente&#46; Los pacientes que presentan este cuadro tienen una mayor activaci&#243;n del complemento y unas cifras de C3a superiores a las de los pacientes que no lo presentan<span class="elsevierStyleSup">21</span>&#46; Cuanto menos biocompatible es la membrana de di&#225;lisis&#44; mayor es la activaci&#243;n del complemento&#44; de tal manera que estas reacciones son m&#225;s frecuentes con las membranas celul&#243;sicas que con las sint&#233;ticas&#46; El AN69 induce una m&#237;nima activaci&#243;n del complemento&#44; mientras que la polisulfona s&#237; lo activa&#44; pero las anafilotoxinas quedan adsorbidas por la membrana y su efecto sist&#233;mico es m&#237;nimo&#46;</p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">Otras reacciones</span></p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara">En 2001 se produjo una inesperada alta mortalidad de pacientes en HD&#44; durante un per&#237;odo concreto de tiempo&#44; localizada geogr&#225;ficamente en Croacia&#44; Espa&#241;a y Estados Unidos&#44; asociada al uso de dializadores de la serie Althane &#40;Baxter&#41;&#46; La muerte se produc&#237;a durante la sesi&#243;n de HD o brevemente despu&#233;s de su finalizaci&#243;n y no cumpl&#237;a exactamente los criterios que definen las reacciones de hipersensibilidad&#46; Se denomin&#243; como <span class="elsevierStyleItalic">perfluorocarbon syndrome</span>&#46; En una investigaci&#243;n posterior se demostr&#243; que el l&#237;quido PF-5070&#44; que se utilizaba como un test para detectar fugas capilares durante la fabricaci&#243;n del dializador&#44; era el causante de esta epidemia de mortalidad&#46; Se concluy&#243; que el PF-5070 es un compuesto altamente t&#243;xico cuando se administra por v&#237;a intravenosa&#44; debido a sus propiedades emulsionantes&#46; Su uso o el de cualquiera de los compuestos de fluorocarbono l&#237;quido debe evitarse en dispositivos m&#233;dicos en contacto con la sangre y en particular en la fabricaci&#243;n de dializador<span class="elsevierStyleSup">24</span>&#46;</p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">HIPERSENSIBILIDAD A MEMBRANAS SINT&#201;TICAS</span></p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara">En el presente n&#250;mero de la revista se publica una serie de casos cl&#237;nicos de dos hospitales de Madrid&#44; en la que se describe la aparici&#243;n de reacciones de hipersensibilidad a membranas sint&#233;ticas de di&#225;lisis en siete pacientes&#46; Trataremos de dilucidar en este comentario editorial en qu&#233; tipo de reacci&#243;n se pueden encuadrar&#44; qu&#233; tienen en com&#250;n todos los casos y si estamos frente a un nuevo problema o se trata de una mera coincidencia temporal de casos esperables&#44; habida cuenta del alto n&#250;mero de pacientes prevalentes en nuestras unidades de HD y la alta proporci&#243;n de membranas sint&#233;ticas que se utilizan en la actualidad&#46; S&#225;nchez-Villanueva et al&#46;<span class="elsevierStyleSup">25</span> comunican 6 casos&#58; el primer paciente sufre una reacci&#243;n con un dializador de poliamida&#44; que combina pol&#237;meros de poliamida&#44; poliariletersulfona y polivinilpirrolidona &#40;PVP&#41; y con otro dializador de polinefrona &#40;polietersulfona de &#250;ltima generaci&#243;n&#41;&#46; El segundo presenta reacciones con polinefrona y helixona &#40;polisulfona&#41;&#46; Los otros cuatro sufren reacciones con un dializador de helixona&#46; Mart&#237;n-Navarro et al&#46;<span class="elsevierStyleSup">26</span> comunican un caso de un paciente que sufre reacciones de hipersensibilidad con diferentes membranas sint&#233;ticas&#58; poliamida&#44; helixona y PMMA&#46; Con un dializador de PAN &#40;AN69&#41; sufre urticaria y eosinofilia y con PAN &#40;AN69&#41; en placas&#44; sin PVP&#44; solamente tiene inestabilidad hemodin&#225;mica&#46; En todos los casos las reacciones desaparecen cuando se cambia a un dializador de triacetato de celulosa&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">&#160;</span></p><p class="elsevierStylePara"><span class="elsevierStyleBold">&#191;Qu&#233; tipo de reacci&#243;n presentan estos pacientes&#63;</span></p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara">La mayor&#237;a de los casos cumple algunos criterios de una reacci&#243;n de tipo A y otros de una reacci&#243;n de tipo B&#46; El momento de su aparici&#243;n es muy variable&#58; desde cinco minutos despu&#233;s de iniciar la di&#225;lisis<span class="elsevierStyleSup">26</span> hasta una hora antes de finalizarla<span class="elsevierStyleSup">25</span>&#46; La mayor&#237;a de los pacientes no presenta signos de angioedema y lo que predomina es la sintomatolog&#237;a respiratoria &#40;disnea y descenso de la saturaci&#243;n de ox&#237;geno&#41;&#44; as&#237; como la hipotensi&#243;n arterial&#46; Adem&#225;s&#44; en alg&#250;n caso no fue preciso desconectar al paciente&#44; ya que el cuadro cl&#237;nico tendi&#243; a mejorar a medida que transcurr&#237;a la HD&#44; aunque&#44; en general&#44; se trat&#243; de cuadros graves&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">&#160;</span></p><p class="elsevierStylePara"><span class="elsevierStyleBold">&#191;Qu&#233; tienen en com&#250;n estos casos&#63;</span></p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara">Todos se producen con membranas sint&#233;ticas y no aparecen cuando se utiliza una membrana celul&#243;sica sustituida &#40;triacetato de celulosa&#41;&#46; Da la impresi&#243;n de que los pacientes son sensibles a membranas sint&#233;ticas de diferente composici&#243;n&#44; aunque en la mayor&#237;a de los casos una de las membranas usadas es una polisulfona y casi todas contienen PVP&#44; que se utiliza para hidrofilizar la membrana de polisulfona y otras membranas sint&#233;ticas&#46; En todos los casos se descartaron otras causas de reacci&#243;n de hipersensibilidad en HD&#44; como la alergia al hierro intravenoso&#44; a la heparina o al l&#225;tex&#46; Adem&#225;s&#44; en ambas unidades se utiliza agua ultrapura&#44; con lo que se puede descartar la reacci&#243;n a pir&#243;genos que aparece al principio de la di&#225;lisis por retrofiltraci&#243;n&#44; a trav&#233;s de la membrana&#44; de un l&#237;quido de di&#225;lisis contaminado al compartimento sangu&#237;neo<span class="elsevierStyleSup">27</span>&#46; En cuanto al m&#233;todo de esterilizaci&#243;n de los dializadores&#44; en ning&#250;n caso se utiliz&#243; el OE&#44; siendo la radiaci&#243;n gamma y el vapor los empleados&#46; Aunque se ha publicado que el agente esterilizante puede afectar a la biocompatibilidad de la membrana<span class="elsevierStyleSup">28</span>&#44; no se ha descrito que la radiaci&#243;n gamma o el vapor produzcan reacciones de hipersensibilidad&#46;</p><p class="elsevierStylePara">En todos los pacientes&#44; la sustituci&#243;n de la membrana sint&#233;tica por otra celul&#243;sica hizo que las reacciones no volvieran a aparecer&#46; En principio&#44; cualquier membrana celul&#243;sica&#44; por su capacidad para generar anafilotoxinas&#44; podr&#237;a producir una reacci&#243;n de tipo B&#46; Sin embargo&#44; la celulosa sustituida es m&#225;s biocompatible que otras celul&#243;sicas&#44; y concretamente el triacetato se comporta como un dializador de alta permeabilidad&#44; con menor capacidad de activar el complemento y con una gran biocompatibilidad&#46; Se ha descrito que la polisulfona&#44; en comparaci&#243;n con el triacetato de celulosa&#44; produce una mayor activaci&#243;n del receptor de membrana de las plaquetas GPIIb&#47;IIIa<span class="elsevierStyleSup">29</span>&#46; Esta glicoprote&#237;na es el receptor para el fibrin&#243;geno que media en la agregaci&#243;n y la adhesi&#243;n plaquetaria y podr&#237;a ser un marcador de biocompatibilidad&#46; Adem&#225;s&#44; en un estudio reciente se ha demostrado por prote&#243;mica que ambas membranas permiten la adsorci&#243;n de diferentes prote&#237;nas plasm&#225;ticas&#46; Por otra parte&#44; el triacetato de celulosa es capaz de adsorber una gran cantidad de alb&#250;mina y apolipoprote&#237;nas&#44; lo que incrementar&#237;a su biocompatibilidad y reducir&#237;a la agregaci&#243;n plaquetaria&#46; La polisulfona adsorbe m&#225;s prote&#237;nas que participan en la interacci&#243;n sangre-membrana&#44; como la ficolina-2&#44; fragmentos de fibrin&#243;geno y prote&#237;nas procedentes de los hemat&#237;es &#40;como la anhidrasa carb&#243;nica y la hemoglobina&#41;&#44; que podr&#237;an ser un signo de <span class="elsevierStyleItalic">shear stress</span> de los hemat&#237;es y de hem&#243;lisis parcial<span class="elsevierStyleSup">30</span>&#46; La ficolina-2&#44; que participa en la v&#237;a del complemento dependiente de la lectina&#44; tambi&#233;n se ha encontrado adsorbida en la membrana de polisulfona por otros autores<span class="elsevierStyleSup">31&#44;32</span>&#44; lo que sugiere que esta membrana podr&#237;a&#44; en algunos casos concretos&#44; activar el complemento y&#44; en teor&#237;a&#44; dar lugar a una reacci&#243;n de tipo B&#46; Sin embargo&#44; su gran capacidad adsortiva de las fracciones del complemento hace que una reacci&#243;n por anafilotoxinas sea poco frecuente&#46; Por &#250;ltimo&#44; la PVP que lleva la mayor&#237;a de las membranas de polisulfona &#40;y no la de triacetato de celulosa&#41; podr&#237;a desempe&#241;ar un papel&#46; La PVP es un conocido al&#233;rgeno que se utiliza para hidrolizar la membrana e inhibir su interacci&#243;n con las prote&#237;nas plasm&#225;ticas y las plaquetas&#46; Su liberaci&#243;n al torrente circulatorio durante la HD puede dar lugar a una reacci&#243;n anafil&#225;ctica grave&#46; Recientemente&#44; se ha descrito un caso de reacciones de tipo A&#44; con IgE elevad&#237;sima&#44; en un paciente tratado con dializadores de polisulfona m&#225;s PVP de diferentes fabricantes y con diferentes m&#233;todos de esterilizaci&#243;n&#44; que desaparecieron cuando se cambi&#243; el dializador por otro de diacetato de celulosa<span class="elsevierStyleSup">33</span>&#46; El caso del paciente de Mart&#237;n-Navarro et al&#46; con un dializador de AN69 con PVP sufre urticaria y eosinofilia y&#44; cuando se cambia a un dializador en placas de AN69 sin PVP&#44; los s&#237;ntomas desaparecen&#46;</p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">&#191;Estamos ante una &#171;epidemia&#187; o es una mera coincidencia temporal de casos&#63;</span></p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara">En el estudio de Simon et al&#46;&#44; el riesgo relativo de que ocurriera una reacci&#243;n de hipersensibilidad fue de 10 a 20 veces m&#225;s alto con membranas sint&#233;ticas que con membranas celul&#243;sicas&#46; La prevalencia de una reacci&#243;n severa fue 0&#44;25&#160;&#37; en el total de la poblaci&#243;n en di&#225;lisis&#44; de 0&#44;5&#160;&#37; en pacientes tratados con membranas sint&#233;ticas&#44; de 1&#44;1&#160;&#37; en aquellos con AN69 y de 4&#44;9&#160;&#37; en los tratados con AN69 e IECA<span class="elsevierStyleSup">4</span>&#46; Esto quiere decir que las reacciones a membranas sint&#233;ticas no son una rareza&#46; El hecho de que en uno de los centros de Madrid el mayor n&#250;mero de casos se concentre en dos a&#241;os puede ser casual&#44; ya que no se comunicaron m&#225;s casos en esa fecha y la Agencia Espa&#241;ola del Medicamento no hab&#237;a recibido otras notificaciones&#44; como indican los propios autores&#46; Se han descrito casos de reacci&#243;n severa a la polisulfona&#44; con y sin PVP<span class="elsevierStyleSup">33&#44;34</span>&#44; tanto de alta como de mediana permeabilidad<span class="elsevierStyleSup">35&#44;36</span>&#44; con filtros de polietersulfona de muy alta permeabilidad &#40;PUREMA<span class="elsevierStyleSup">&#174;</span>&#41;<span class="elsevierStyleSup">37</span> y con un dializador de polisulfona de un fabricante s&#237; y de otro no<span class="elsevierStyleSup">38</span>&#46; Tambi&#233;n se han descrito reacciones cruzadas entre polisulfona&#44; PMMA y policarbonato<span class="elsevierStyleSup">39</span> en un mismo paciente&#46;</p><p class="elsevierStylePara">A pesar de las sustanciales mejoras en la biocompatibilidad de las membranas&#44; los filtros&#44; los m&#233;todos de esterilizaci&#243;n&#44; etc&#46;&#44; la exposici&#243;n repetida de la sangre a sustancias extra&#241;as podr&#237;a producir una sensibilizaci&#243;n en pacientes predispuestos y favorecer la aparici&#243;n de reacciones de hipersensibilidad&#46;</p><p class="elsevierStylePara">En resumen&#44; se ha descrito la aparici&#243;n de reacciones de hipersensibilidad inespec&#237;ficas con membranas sint&#233;ticas&#44; m&#225;s frecuentemente con polisulfona &#40;que son tambi&#233;n las m&#225;s utilizadas&#41;&#44; durante un per&#237;odo de tiempo concreto y agrupadas en algunos hospitales&#44; cuya etiolog&#237;a no ha sido claramente definida y que est&#225; a&#250;n pendiente de tipificar&#46;</p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">Conflictos de inter&#233;s</span></p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara">Los autores declaran los siguientes conflictos de inter&#233;s potenciales&#58;</p><p class="elsevierStylePara">Alejandro Mart&#237;n-Malo ha recibido honorarios de Abbie&#44; Amgen&#44; Bellco&#44; Fresenius Medical Care&#44; Gambro y Shire y M&#170; Antonia &#193;lvarez de Lara ha recibido honorarios de Abbie&#44; Amgen&#44; Bellco&#44; Fresenius Medical&#46;</p><p class="elsevierStylePara">&#160;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">CONCEPTOS CLAVE</span></p><p class="elsevierStylePara">&#160;</p><li>Las reacciones adversas en di&#225;lisis son el resultado de la interacci&#243;n entre la sangre y los diferentes materiales que forman el dializador y los dem&#225;s componentes del circuito extracorp&#243;reo&#46; Pueden ser de tipo A o de hipersensibilidad &#40;anafil&#225;cticas o anafilactoides&#41; y de tipo B o inespec&#237;ficas &#40;generalmente mediadas por la activaci&#243;n del complemento&#41;&#46;</li><li>Pueden aparecer reacciones tanto con las membranas celul&#243;sicas como con las sint&#233;ticas&#44; siendo hoy en d&#237;a m&#225;s frecuentes con estas &#250;ltimas&#46;</li><li>Es posible que el uso de PVP&#44; una sustancia muy alerg&#233;nica que se usa para hidrofilizar algunas membranas&#44; aumente la probabilidad de sufrir una reacci&#243;n de hipersensibilidad&#46;</li><li>En pacientes que sufran una reacci&#243;n de hipersensibilidad en di&#225;lisis hay que descartar otras causas como el l&#225;tex&#44; el hierro intravenoso&#44; la heparina y el formaldeh&#237;do&#46; </li>"
    "pdfFichero" => "P1-E577-S4787-A12682.pdf"
    "tienePdf" => true
    "bibliografia" => array:2 [
      "titulo" => "Bibliograf&#237;a"
      "seccion" => array:1 [
        0 => array:1 [
          "bibliografiaReferencia" => array:39 [
            0 => array:3 [
              "identificador" => "bib1"
              "etiqueta" => "1"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Poothullil J, Shimizu A, Day RP, Dolovich J. Anaphylaxis from the product(s) of ethylene oxide gas. Ann Intern Med 1975;82:58-60. <a href="http://www.ncbi.nlm.nih.gov/pubmed/1235763" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            1 => array:3 [
              "identificador" => "bib2"
              "etiqueta" => "2"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Daugirdas JT, Ing TS, Roxe DM, Ivanovich PT, Krumlovsky F, Popli S, et al. Severe anaphylactoid reactions to cuprammonium cellulose hemodialyzers. Ann Intern Med 1985;145:489-94."
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            2 => array:3 [
              "identificador" => "bib3"
              "etiqueta" => "3"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Nicholls AJ. Hypersensitivity to hemodialysis: the United Kingdom experience. Artif Organs 1987;11(2):87-9. <a href="http://www.ncbi.nlm.nih.gov/pubmed/3593049" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            3 => array:3 [
              "identificador" => "bib4"
              "etiqueta" => "4"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Simon P, Potier J, Thebaud HE. Risk factors for acute hypersensitivity reactions in hemodialysis. Nephrologie 1996;17:163-70. <a href="http://www.ncbi.nlm.nih.gov/pubmed/9064565" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            4 => array:3 [
              "identificador" => "bib5"
              "etiqueta" => "5"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Ebo DG, Bosman JL, Couttenye MM, Stevens WJ. Haemodialysis-associated anaphylactic and anaphylactoid reactions. Allergy 2006;61:211-20. <a href="http://www.ncbi.nlm.nih.gov/pubmed/16409199" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            5 => array:3 [
              "identificador" => "bib6"
              "etiqueta" => "6"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Daugirdas JT, Ing TS. First-use reactions during hemodialysis: A definition of subtypes. Kidney Int Suppl 1988;24:S37-43. <a href="http://www.ncbi.nlm.nih.gov/pubmed/2966256" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            6 => array:3 [
              "identificador" => "bib7"
              "etiqueta" => "7"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Lemke HD, Heidland A, Schaefer RM. Hypersensitivity reactions during haemodialysis: role of complement fragments and ethylene oxide antibodies. Nephrol Dial Transplant 1990;5:264-9. <a href="http://www.ncbi.nlm.nih.gov/pubmed/2113222" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            7 => array:3 [
              "identificador" => "bib8"
              "etiqueta" => "8"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Pegues DA, Beck-Sague CM, Woollen SW, Greenspan B, Burns SM, Bland LA, et al. Anaphylactoid reactions associated with reuse of hollow-fiber hemodialyzers and ACE inhibitors. Kidney Int 1992;42:1232-7. <a href="http://www.ncbi.nlm.nih.gov/pubmed/1453608" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:1 [
                        "itemHostRev" => array:3 [
                          "pii" => "S0022347608002242"
                          "estado" => "S300"
                          "issn" => "00223476"
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            8 => array:3 [
              "identificador" => "bib9"
              "etiqueta" => "9"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Centers for Disease Control (CDC). Acute allergic reactions associated with reprocessed hemodialyzers¿United States, 1989-1990. MMWR Morb Mortal Wkly Rep 1991;40(9):147, 153-4."
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            9 => array:3 [
              "identificador" => "bib10"
              "etiqueta" => "10"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Tielemans C, Madhoun P, Lenaers M, Schandene L, Goldman M, Vanherweghem JL. Anaphylactoid reactions during hemodialysis on AN69 membranes in patients receiving ACE inhibitors. Kidney Int 1990;38:982-4. <a href="http://www.ncbi.nlm.nih.gov/pubmed/2266684" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            10 => array:3 [
              "identificador" => "bib11"
              "etiqueta" => "11"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Alvarez-Lara MA, Martín-Malo A, Espinosa M, Castillo D, Aljama P. ACE inhibitors and anaphilactoid reactions during AN-69 membrane dialysis. Lancet 1991;337(8737):370-1. <a href="http://www.ncbi.nlm.nih.gov/pubmed/1671272" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            11 => array:3 [
              "identificador" => "bib12"
              "etiqueta" => "12"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Schulman G, Hakim R, Arias R, Silverberg M, Kaplan AP, Arbeit L. Bradykinin generation by dialysis membranes: possible role in anaphylactic reaction. J Am Soc Nephrol 1993;3:1563-9. <a href="http://www.ncbi.nlm.nih.gov/pubmed/8507811" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            12 => array:3 [
              "identificador" => "bib13"
              "etiqueta" => "13"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Verresen L, Fink E, Lemke HD, Vanrenterghem Y. Bradykinin is a mediator of anaphylactoid reactions during hemodialysis with AN69 membranes. Kidney Int 1994;45:1497-503. <a href="http://www.ncbi.nlm.nih.gov/pubmed/8072263" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            13 => array:3 [
              "identificador" => "bib14"
              "etiqueta" => "14"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Maheut H, Lacour F. Using AN69 ST membrane: a dialysis centre experience. Nephrol Dial Transplant 2001;16:1519-20. <a href="http://www.ncbi.nlm.nih.gov/pubmed/11427664" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            14 => array:3 [
              "identificador" => "bib15"
              "etiqueta" => "15"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Maurice F, Rivory JP, Larsson PH, Johansson SG, Bousquet J. Anaphylactic shock caused by formaldehyde in a patient undergoing long-term hemodialysis. J Allergy Clin Immunol 1986;77:594-7. <a href="http://www.ncbi.nlm.nih.gov/pubmed/3958388" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            15 => array:3 [
              "identificador" => "bib16"
              "etiqueta" => "16"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Kalpaklioglu AF, Aydin G. Prevalence of latex sensitivity among patients with chronic renal failure: a new risk group? Artif Organs 1999;23:139-42. <a href="http://www.ncbi.nlm.nih.gov/pubmed/10027882" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            16 => array:3 [
              "identificador" => "bib17"
              "etiqueta" => "17"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Ebo DG, Haine SE, Hagendorens MM, Bridts CH, Conraads VM, Vorlat A, et al. Hypersensitivity to nadroparin calcium: case report and review of the literature. Clin Drug Investig 2004;24:421-6. <a href="http://www.ncbi.nlm.nih.gov/pubmed/17516728" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            17 => array:3 [
              "identificador" => "bib18"
              "etiqueta" => "18"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Berkun Y, Aviv YS, Schwartz LB, Shalit M. Heparin-induced recurrent anaphylaxis. Clin Exp Allergy 2004;34(12):1916-8. <a href="http://www.ncbi.nlm.nih.gov/pubmed/15663568" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            18 => array:3 [
              "identificador" => "bib19"
              "etiqueta" => "19"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Michael B, Coyne DW, Fishbane S, Folkert V, Lynn R, Nissenson AR, et al. Sodium ferric gluconate complex in hemodialysis patients: adverse reactions compared to placebo and iron dextran. Kidney Int 2002;61:1830-9. <a href="http://www.ncbi.nlm.nih.gov/pubmed/11967034" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            19 => array:3 [
              "identificador" => "bib20"
              "etiqueta" => "20"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Craddock PR, Fehr J, Dalmasso AP, Brighan KL, Jacob HS. Hemodialysis leukopenia. Pulmonary vascular leukostasis resulting from complement activation by dialyzer cellophane membranes. J Clin Invest 1977;59:879-88. <a href="http://www.ncbi.nlm.nih.gov/pubmed/856872" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            20 => array:3 [
              "identificador" => "bib21"
              "etiqueta" => "21"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Hakim RM, Breillatt J, Lazarus JM, Port FK. Complement activation and hypersensitivity reactions to dialysis membranes. N Engl J Med 1984;311:878-82. <a href="http://www.ncbi.nlm.nih.gov/pubmed/6332276" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            21 => array:3 [
              "identificador" => "bib22"
              "etiqueta" => "22"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Aljama P, Martín-Malo A, Castillo D, Velasco F, Torres A, Pérez R, et al. Anaphylatoxin C5a generation and dialysis-induced leukopenia with different hemodialyzer membranes. Blood Purif 1986;4:88-92. <a href="http://www.ncbi.nlm.nih.gov/pubmed/3730166" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            22 => array:3 [
              "identificador" => "bib23"
              "etiqueta" => "23"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Martin-Malo A, Castillo D, Castro M, Pérez R, Rios A, Jaraba M, et al. Biocompatibility of dialysis membranes: a comparative study. Nephrol Dial Transplant 1991;6 Suppl 2:55-8."
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            23 => array:3 [
              "identificador" => "bib24"
              "etiqueta" => "24"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Canaud B, Aljama P, Tielemans C, Gasparovic V, Gutierrez A, Locatelli F. Pathochemical toxicity of perfluorocarbon-5070, a liquid test performance fluid previously used in dialyzer manufacturing, confirmed in animal experiment. J Am Soc Nephrol 2005;16:1819-2."
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            24 => array:3 [
              "identificador" => "bib25"
              "etiqueta" => "25"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Sánchez-Villanueva RJ, González E, Quirce S, Díaz R, Álvarez L, Menéndez D, et al. Reacciones de hipersensibilidad a membranas sintéticas de hemodiálisis. Nefrologia 2014;34(4):520-5. <a href="http://www.ncbi.nlm.nih.gov/pubmed/25036066" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            25 => array:3 [
              "identificador" => "bib26"
              "etiqueta" => "26"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Martín-Navarro JA, Gutiérrez-Sánchez MJ, Petkov-Stoyanov V. Hipersensibilidad a membranas sintéticas de hemodiálisis. Nefrologia 2014;34(6):XXX-XX."
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            26 => array:3 [
              "identificador" => "bib27"
              "etiqueta" => "27"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Bigazzi R, Atti M, Baldari G. High-permeable membranes and hypersensitivity-like reactions: role of dialysis fluid contamination. Blood Purif 1990;8(4):190-8. <a href="http://www.ncbi.nlm.nih.gov/pubmed/2085427" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            27 => array:3 [
              "identificador" => "bib28"
              "etiqueta" => "28"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Muller TF, Seitz M, Eckle I, Lange H, Kolb G. Biocompatibility differences with respect to the dialyzer sterilization method. Nephron 1998;78:139-42. <a href="http://www.ncbi.nlm.nih.gov/pubmed/9496728" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            28 => array:3 [
              "identificador" => "bib29"
              "etiqueta" => "29"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Kuragano T, Kuno T, Takahashi Y, Yamamoto C, Nagura Y, Takahashi S, et al. Comparison of the effects of cellulose triacetate and polysulfone membrane on GPIIb/IIIa and platelet activation. Blood Purif 2003;21:176-82. <a href="http://www.ncbi.nlm.nih.gov/pubmed/12601261" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            29 => array:3 [
              "identificador" => "bib30"
              "etiqueta" => "30"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Urbani A, Sirolli V, Lupisella S, Levi-Mortera S, Pavone B, Pieroni L, et al. Proteomic investigations on the effect of different membrane materials on blood protein adsorption during haemodialysis. Blood Transfus 2012;10 Suppl 2:s101-12. <a href="http://www.ncbi.nlm.nih.gov/pubmed/22890260" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            30 => array:3 [
              "identificador" => "bib31"
              "etiqueta" => "31"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Mares J, Thongboonkerd V, Tuma Z, Moravec J, Matejovic M. Specific adsorption of some complement activation proteins to polysulfone dialysis membranes during hemodialysis. Kidney Int 2009;76:404-13. <a href="http://www.ncbi.nlm.nih.gov/pubmed/19421191" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            31 => array:3 [
              "identificador" => "bib32"
              "etiqueta" => "32"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Mares J, Richtrova P, Hricinova A, Tuma Z, Moravec J, Lysak D, et al. Proteomic profiling of blood-dialyzer interactome reveals involvement of lectin complement pathway in hemodialysis-induced inflammatory response. Proteomics Clin Appl 2010;4:829-38. <a href="http://www.ncbi.nlm.nih.gov/pubmed/21137026" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            32 => array:3 [
              "identificador" => "bib33"
              "etiqueta" => "33"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Marques ID, Pinheiro KF, de Freitas do Carmo LP, Costa MC, Abensur H. Anaphylactic reaction induced by a polysulfone/polyvinylpyrrolidone membrane in the 10th session of hemodialysis with the same dialyzer. Hemodial Int 2011;15(3):399-403. <a href="http://www.ncbi.nlm.nih.gov/pubmed/21624039" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            33 => array:3 [
              "identificador" => "bib34"
              "etiqueta" => "34"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Arenas MD, Gil MT, Carreton MA, Moledous A, Albiach B. Eventos adversos a polisulfona durante la sesión de hemodiálisis. Nefrologia 2007;27:638-42. <a href="http://www.ncbi.nlm.nih.gov/pubmed/18045043" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            34 => array:3 [
              "identificador" => "bib35"
              "etiqueta" => "35"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Huang WH, Lee YY, Shih LC. Delayed near-fatal anaphylactic reaction induced by the F10-HPS polysulphone haemodialyser. Nephrol Dial Transplant 2008;23(1):423-4. <a href="http://www.ncbi.nlm.nih.gov/pubmed/17893104" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            35 => array:3 [
              "identificador" => "bib36"
              "etiqueta" => "36"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Arenas MD, Niveiro E, Moledous A, Gil MT, Albiach B, Carretón MA. Fatal acute systemic hypersensitivity reaction during haemodialysis. Nephrol Dial Transplant 2006;21(10):2966-70. <a href="http://www.ncbi.nlm.nih.gov/pubmed/16854851" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            36 => array:3 [
              "identificador" => "bib37"
              "etiqueta" => "37"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Heegard KD, Tilley MA, Stewart IJ, Edgecombe HP, Lundy JB, Renz EM, et al. Anaphylactoid reaction during first hemofiltration with a PUREMA polysulfone membrane. Int J Artif Organs 2013;36(5):363-6. <a href="http://www.ncbi.nlm.nih.gov/pubmed/23420464" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            37 => array:3 [
              "identificador" => "bib38"
              "etiqueta" => "38"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Ohashi N, Yonemura K, Goto T, Suzuki H, Fujigaki Y, Yamamoto T, et al. A case of anaphylactoid shock induced by the BS polysulfone hemodialyzer but not by the F8-HPS polysulfone hemodialyzer. Clin Nephrol 2003;60:214-7. <a href="http://www.ncbi.nlm.nih.gov/pubmed/14524587" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
            38 => array:3 [
              "identificador" => "bib39"
              "etiqueta" => "39"
              "referencia" => array:1 [
                0 => array:3 [
                  "referenciaCompleta" => "Tanoue N, Nagano K, Matsumura H. Use of a light-polymerized composite removable partial denture base for a patient hypersensitive to poly(methyl methacrylate), polysulfone, and polycarbonate: a clinical report. J Prosthet Dent 2005;93:17-20. <a href="http://www.ncbi.nlm.nih.gov/pubmed/15623992" target="_blank">[Pubmed]</a>"
                  "contribucion" => array:1 [
                    0 => null
                  ]
                  "host" => array:1 [
                    0 => null
                  ]
                ]
              ]
            ]
          ]
        ]
      ]
    ]
  ]
  "idiomaDefecto" => "es"
  "url" => "/02116995/0000003400000006/v0_201502091330/X0211699514054894/v0_201502091331/es/main.assets"
  "Apartado" => array:4 [
    "identificador" => "35355"
    "tipo" => "SECCION"
    "es" => array:2 [
      "titulo" => "Comentarios Editoriales"
      "idiomaDefecto" => true
    ]
    "idiomaDefecto" => "es"
  ]
  "PDF" => "https://static.elsevier.es/multimedia/02116995/0000003400000006/v0_201502091330/X0211699514054894/v0_201502091331/es/P1-E577-S4787-A12682.pdf?idApp=UINPBA000064&text.app=https://revistanefrologia.com/"
  "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699514054894?idApp=UINPBA000064"
]
Compartir
Información de la revista
Vol. 34. Núm. 6.noviembre 2014
Páginas 693-810
Vol. 34. Núm. 6.noviembre 2014
Páginas 693-810
Acceso a texto completo
Hipersensibilidad a membranas sintéticas de hemodiálisis, ¿un problema emergente?
Hypersensitivity reactions to synthetic haemodialysis membranes ¿ an emerging issue?
Visitas
37423
M. Antonia Álvarez-de Laraa, Alejandro Martín-Maloa
a Servicio de Nefrología, Hospital Universitario Reina Sofía, Córdoba,
Contenido relacionado
Nefrología al día. 2012;7:010.3265/Nefrologia.2010.pub1.ed80.chapter2810
Alejandro Martín Malo, ALM de Francisco
Este artículo ha recibido
Información del artículo
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Texto completo

INTRODUCCIÓN

 

Las reacciones de tipo anafiláctico o anafilactoide que aparecen durante la hemodiálisis (HD) se conocen desde 19751. Desde entonces se han publicado numerosos artículos en los que se comunican este tipo de reacciones, aunque la mayoría de ellos con pocos casos y con diferentes membranas, agentes esterilizantes, medicación administrada durante la diálisis, grado de pureza del agua, etc. No hay estudios epidemiológicos prospectivos bien diseñados que nos permitan conocer la incidencia exacta y el alcance de este tipo de reacciones. En un estudio realizado en 1985, Daugirdas comunicó 21 reacciones severas, con un caso fatal, en 260 000 sesiones de HD2. En 1987 se investigó la prevalencia de reacciones de hipersensibilidad «de primer uso» en Reino Unido y se encontró que 1 de cada 20 a 50 pacientes era susceptible de presentar una reacción anafilactoide con un nuevo dializador, lo cual indica la magnitud del problema3. Estas reacciones no se relacionaban con una membrana o tipo de dializador o con una técnica de diálisis determinados. Años después, otro estudio analizó la incidencia de reacciones en 1536 pacientes de 30 centros de diálisis (122 694 sesiones) y se observó una incidencia anual de 0,17 por 1000 sesiones con membranas celulósicas y 4,2 por 1000 sesiones con membranas sintéticas4. Esto indica que, aunque no son extremadamente frecuentes, las reacciones aparecen de vez en cuando en todas las unidades de diálisis y que, aparentemente, es más frecuente que estén asociadas al uso de membranas sintéticas.

 

TIPOS DE REACCIÓN

 

Las reacciones que aparecen durante la HD son el resultado de una respuesta inmunoalérgica por parte del paciente tras la exposición a sustancias extrañas al organismo que están presentes en el circuito extracorpóreo y/o por la respuesta inducida por la interacción de la sangre con la membrana del dializador5. Pueden ser de dos tipos6.

 

Tipo A o reacciones de hipersensibilidad

 

Se presentan normalmente en los primeros minutos de la diálisis, aunque pueden aparecer hasta 30 minutos después del inicio. Se caracterizan por urticaria, tos, rinorrea, lagrimeo, calambres abdominales, prurito, sensación de quemazón, angioedema, disnea e incluso colapso circulatorio y muerte. Son reacciones graves que obligan a parar inmediatamente la diálisis, siendo aconsejable no retornar la sangre del circuito extracorpóreo. Estas reacciones de tipo A pueden ser, a su vez, anafilácticas cuando están mediadas por la IgE o anafilactoides si no están mediadas por la IgE.

Las más características se han descrito en relación con el óxido de etileno (OE), el reúso de dializadores y por la combinación de membranas de poliacrilonitrilo (PAN, AN69) e inhibidores de la enzima conversora de la angiotensina (IECA).

La clásica reacción al OE se daba exclusivamente durante el primer uso de un dializador esterilizado con OE que no había sido cebado adecuadamente. En la mayoría de los casos existía una IgE frente al OE detectable en el plasma de los pacientes que sufrían este tipo de reacciones1,7. En la actualidad este agente esterilizante ya no se utiliza en los dializadores y ha sido sustituido por radiación gamma o vapor. De todos modos, ante una reacción en diálisis hay que tener presente que el OE aún se utiliza para esterilizar algunas agujas, jeringas y líneas de diálisis. También se han descrito reacciones anafilactoides cuando se reutilizan dializadores de polisulfona y acetato de celulosa8, así como un aumento del riesgo de aparición de reacciones de hipersensibilidad cuando se usaba hipoclorito sódico o peróxido de hidrógeno para lavar el compartimento sanguíneo de los dializadores9.

En los años noventa se describió la aparición de reacciones anafilactoides en pacientes dializados con membrana de AN69 (PAN) y que recibían, al mismo tiempo, tratamiento con IECA4,8,10,11. Eran reacciones de tipo A, por el momento de su aparición (en los primeros minutos de la diálisis) y por la sintomatología de los pacientes. El mediador de estas reacciones es la bradikinina. La superficie de la membrana de AN69 tiene una carga eléctrica fuertemente negativa capaz de activar el sistema de contacto e inducir la producción de factor de Hageman. Este convierte la prekalicreína en kalicreína, que actúa sobre el kininógeno para liberar bradikinina. Los IECA, que inhiben la degradación de la bradikinina, hacen que esta se acumule en sangre hasta un nivel de 20 a 30 veces superior al normal, facilitando la aparición de los síntomas de anafilaxia12,13. El recubrimiento de la superficie de la membrana de AN69 con un polímero biocompatible (SPAN, AN69 ST) proporciona una neutralización parcial de la electronegatividad y una reducción en la producción de bradikinina. Con esta modificación en la membrana se ha podido dializar con AN69 ST a pacientes tratados con IECA y con historia de reacción anafilactoide durante la HD con AN69, sin que presentasen ningún problema14.

Por último, también pueden causar reacciones de hipersensibilidad otros compuestos utilizados en las unidades de diálisis como el formaldehído15, el látex16, la heparina17,18 y el hierro intravenoso19. Esto hace que, con frecuencia, ante una reacción de tipo A durante la HD sea difícil encontrar el agente causal.

 

Tipo B o inespecíficas

 

Son más frecuentes y menos graves que las de tipo A. Los síntomas son dolor torácico, disnea, náuseas, vómitos e hipotensión. Son más tardías, aparecen a los 15-30 minutos de iniciada la diálisis, aunque pueden hacerlo incluso más tarde, y generalmente se resuelven a medida que transcurre la sesión, sin necesidad de desconectar al paciente. Son debidas a una leucostasis pulmonar secundaria a la activación del complemento por la membrana de diálisis que genera C3a y C5a. Los grupos hidroxilo libres de la membrana de diálisis activan la vía alterna del complemento, generándose las anafilotoxinas C3a y C5a. Esta última se une a los receptores de membrana de los leucocitos produciendo activación, agregación y adherencia de estos al endotelio del capilar pulmonar, lo que da lugar a las ya mencionadas leucostasis pulmonar, leucopenia e hipoxemia20-23. El nadir de la leucopenia se produce a los 15 minutos de iniciada la sesión y posteriormente la cifra de leucocitos en sangre va aumentando y recupera el valor prediálisis a la hora aproximadamente. Los pacientes que presentan este cuadro tienen una mayor activación del complemento y unas cifras de C3a superiores a las de los pacientes que no lo presentan21. Cuanto menos biocompatible es la membrana de diálisis, mayor es la activación del complemento, de tal manera que estas reacciones son más frecuentes con las membranas celulósicas que con las sintéticas. El AN69 induce una mínima activación del complemento, mientras que la polisulfona sí lo activa, pero las anafilotoxinas quedan adsorbidas por la membrana y su efecto sistémico es mínimo.

 

Otras reacciones

 

En 2001 se produjo una inesperada alta mortalidad de pacientes en HD, durante un período concreto de tiempo, localizada geográficamente en Croacia, España y Estados Unidos, asociada al uso de dializadores de la serie Althane (Baxter). La muerte se producía durante la sesión de HD o brevemente después de su finalización y no cumplía exactamente los criterios que definen las reacciones de hipersensibilidad. Se denominó como perfluorocarbon syndrome. En una investigación posterior se demostró que el líquido PF-5070, que se utilizaba como un test para detectar fugas capilares durante la fabricación del dializador, era el causante de esta epidemia de mortalidad. Se concluyó que el PF-5070 es un compuesto altamente tóxico cuando se administra por vía intravenosa, debido a sus propiedades emulsionantes. Su uso o el de cualquiera de los compuestos de fluorocarbono líquido debe evitarse en dispositivos médicos en contacto con la sangre y en particular en la fabricación de dializador24.

 

HIPERSENSIBILIDAD A MEMBRANAS SINTÉTICAS

 

En el presente número de la revista se publica una serie de casos clínicos de dos hospitales de Madrid, en la que se describe la aparición de reacciones de hipersensibilidad a membranas sintéticas de diálisis en siete pacientes. Trataremos de dilucidar en este comentario editorial en qué tipo de reacción se pueden encuadrar, qué tienen en común todos los casos y si estamos frente a un nuevo problema o se trata de una mera coincidencia temporal de casos esperables, habida cuenta del alto número de pacientes prevalentes en nuestras unidades de HD y la alta proporción de membranas sintéticas que se utilizan en la actualidad. Sánchez-Villanueva et al.25 comunican 6 casos: el primer paciente sufre una reacción con un dializador de poliamida, que combina polímeros de poliamida, poliariletersulfona y polivinilpirrolidona (PVP) y con otro dializador de polinefrona (polietersulfona de última generación). El segundo presenta reacciones con polinefrona y helixona (polisulfona). Los otros cuatro sufren reacciones con un dializador de helixona. Martín-Navarro et al.26 comunican un caso de un paciente que sufre reacciones de hipersensibilidad con diferentes membranas sintéticas: poliamida, helixona y PMMA. Con un dializador de PAN (AN69) sufre urticaria y eosinofilia y con PAN (AN69) en placas, sin PVP, solamente tiene inestabilidad hemodinámica. En todos los casos las reacciones desaparecen cuando se cambia a un dializador de triacetato de celulosa.

 

¿Qué tipo de reacción presentan estos pacientes?

 

La mayoría de los casos cumple algunos criterios de una reacción de tipo A y otros de una reacción de tipo B. El momento de su aparición es muy variable: desde cinco minutos después de iniciar la diálisis26 hasta una hora antes de finalizarla25. La mayoría de los pacientes no presenta signos de angioedema y lo que predomina es la sintomatología respiratoria (disnea y descenso de la saturación de oxígeno), así como la hipotensión arterial. Además, en algún caso no fue preciso desconectar al paciente, ya que el cuadro clínico tendió a mejorar a medida que transcurría la HD, aunque, en general, se trató de cuadros graves.

 

¿Qué tienen en común estos casos?

 

Todos se producen con membranas sintéticas y no aparecen cuando se utiliza una membrana celulósica sustituida (triacetato de celulosa). Da la impresión de que los pacientes son sensibles a membranas sintéticas de diferente composición, aunque en la mayoría de los casos una de las membranas usadas es una polisulfona y casi todas contienen PVP, que se utiliza para hidrofilizar la membrana de polisulfona y otras membranas sintéticas. En todos los casos se descartaron otras causas de reacción de hipersensibilidad en HD, como la alergia al hierro intravenoso, a la heparina o al látex. Además, en ambas unidades se utiliza agua ultrapura, con lo que se puede descartar la reacción a pirógenos que aparece al principio de la diálisis por retrofiltración, a través de la membrana, de un líquido de diálisis contaminado al compartimento sanguíneo27. En cuanto al método de esterilización de los dializadores, en ningún caso se utilizó el OE, siendo la radiación gamma y el vapor los empleados. Aunque se ha publicado que el agente esterilizante puede afectar a la biocompatibilidad de la membrana28, no se ha descrito que la radiación gamma o el vapor produzcan reacciones de hipersensibilidad.

En todos los pacientes, la sustitución de la membrana sintética por otra celulósica hizo que las reacciones no volvieran a aparecer. En principio, cualquier membrana celulósica, por su capacidad para generar anafilotoxinas, podría producir una reacción de tipo B. Sin embargo, la celulosa sustituida es más biocompatible que otras celulósicas, y concretamente el triacetato se comporta como un dializador de alta permeabilidad, con menor capacidad de activar el complemento y con una gran biocompatibilidad. Se ha descrito que la polisulfona, en comparación con el triacetato de celulosa, produce una mayor activación del receptor de membrana de las plaquetas GPIIb/IIIa29. Esta glicoproteína es el receptor para el fibrinógeno que media en la agregación y la adhesión plaquetaria y podría ser un marcador de biocompatibilidad. Además, en un estudio reciente se ha demostrado por proteómica que ambas membranas permiten la adsorción de diferentes proteínas plasmáticas. Por otra parte, el triacetato de celulosa es capaz de adsorber una gran cantidad de albúmina y apolipoproteínas, lo que incrementaría su biocompatibilidad y reduciría la agregación plaquetaria. La polisulfona adsorbe más proteínas que participan en la interacción sangre-membrana, como la ficolina-2, fragmentos de fibrinógeno y proteínas procedentes de los hematíes (como la anhidrasa carbónica y la hemoglobina), que podrían ser un signo de shear stress de los hematíes y de hemólisis parcial30. La ficolina-2, que participa en la vía del complemento dependiente de la lectina, también se ha encontrado adsorbida en la membrana de polisulfona por otros autores31,32, lo que sugiere que esta membrana podría, en algunos casos concretos, activar el complemento y, en teoría, dar lugar a una reacción de tipo B. Sin embargo, su gran capacidad adsortiva de las fracciones del complemento hace que una reacción por anafilotoxinas sea poco frecuente. Por último, la PVP que lleva la mayoría de las membranas de polisulfona (y no la de triacetato de celulosa) podría desempeñar un papel. La PVP es un conocido alérgeno que se utiliza para hidrolizar la membrana e inhibir su interacción con las proteínas plasmáticas y las plaquetas. Su liberación al torrente circulatorio durante la HD puede dar lugar a una reacción anafiláctica grave. Recientemente, se ha descrito un caso de reacciones de tipo A, con IgE elevadísima, en un paciente tratado con dializadores de polisulfona más PVP de diferentes fabricantes y con diferentes métodos de esterilización, que desaparecieron cuando se cambió el dializador por otro de diacetato de celulosa33. El caso del paciente de Martín-Navarro et al. con un dializador de AN69 con PVP sufre urticaria y eosinofilia y, cuando se cambia a un dializador en placas de AN69 sin PVP, los síntomas desaparecen.

 

¿Estamos ante una «epidemia» o es una mera coincidencia temporal de casos?

 

En el estudio de Simon et al., el riesgo relativo de que ocurriera una reacción de hipersensibilidad fue de 10 a 20 veces más alto con membranas sintéticas que con membranas celulósicas. La prevalencia de una reacción severa fue 0,25 % en el total de la población en diálisis, de 0,5 % en pacientes tratados con membranas sintéticas, de 1,1 % en aquellos con AN69 y de 4,9 % en los tratados con AN69 e IECA4. Esto quiere decir que las reacciones a membranas sintéticas no son una rareza. El hecho de que en uno de los centros de Madrid el mayor número de casos se concentre en dos años puede ser casual, ya que no se comunicaron más casos en esa fecha y la Agencia Española del Medicamento no había recibido otras notificaciones, como indican los propios autores. Se han descrito casos de reacción severa a la polisulfona, con y sin PVP33,34, tanto de alta como de mediana permeabilidad35,36, con filtros de polietersulfona de muy alta permeabilidad (PUREMA®)37 y con un dializador de polisulfona de un fabricante sí y de otro no38. También se han descrito reacciones cruzadas entre polisulfona, PMMA y policarbonato39 en un mismo paciente.

A pesar de las sustanciales mejoras en la biocompatibilidad de las membranas, los filtros, los métodos de esterilización, etc., la exposición repetida de la sangre a sustancias extrañas podría producir una sensibilización en pacientes predispuestos y favorecer la aparición de reacciones de hipersensibilidad.

En resumen, se ha descrito la aparición de reacciones de hipersensibilidad inespecíficas con membranas sintéticas, más frecuentemente con polisulfona (que son también las más utilizadas), durante un período de tiempo concreto y agrupadas en algunos hospitales, cuya etiología no ha sido claramente definida y que está aún pendiente de tipificar.

 

Conflictos de interés

 

Los autores declaran los siguientes conflictos de interés potenciales:

Alejandro Martín-Malo ha recibido honorarios de Abbie, Amgen, Bellco, Fresenius Medical Care, Gambro y Shire y Mª Antonia Álvarez de Lara ha recibido honorarios de Abbie, Amgen, Bellco, Fresenius Medical.

 

CONCEPTOS CLAVE

 

  • Las reacciones adversas en diálisis son el resultado de la interacción entre la sangre y los diferentes materiales que forman el dializador y los demás componentes del circuito extracorpóreo. Pueden ser de tipo A o de hipersensibilidad (anafilácticas o anafilactoides) y de tipo B o inespecíficas (generalmente mediadas por la activación del complemento).
  • Pueden aparecer reacciones tanto con las membranas celulósicas como con las sintéticas, siendo hoy en día más frecuentes con estas últimas.
  • Es posible que el uso de PVP, una sustancia muy alergénica que se usa para hidrofilizar algunas membranas, aumente la probabilidad de sufrir una reacción de hipersensibilidad.
  • En pacientes que sufran una reacción de hipersensibilidad en diálisis hay que descartar otras causas como el látex, el hierro intravenoso, la heparina y el formaldehído.
  • Bibliografía
    [1]
    Poothullil J, Shimizu A, Day RP, Dolovich J. Anaphylaxis from the product(s) of ethylene oxide gas. Ann Intern Med 1975;82:58-60. [Pubmed]
    [2]
    Daugirdas JT, Ing TS, Roxe DM, Ivanovich PT, Krumlovsky F, Popli S, et al. Severe anaphylactoid reactions to cuprammonium cellulose hemodialyzers. Ann Intern Med 1985;145:489-94.
    [3]
    Nicholls AJ. Hypersensitivity to hemodialysis: the United Kingdom experience. Artif Organs 1987;11(2):87-9. [Pubmed]
    [4]
    Simon P, Potier J, Thebaud HE. Risk factors for acute hypersensitivity reactions in hemodialysis. Nephrologie 1996;17:163-70. [Pubmed]
    [5]
    Ebo DG, Bosman JL, Couttenye MM, Stevens WJ. Haemodialysis-associated anaphylactic and anaphylactoid reactions. Allergy 2006;61:211-20. [Pubmed]
    [6]
    Daugirdas JT, Ing TS. First-use reactions during hemodialysis: A definition of subtypes. Kidney Int Suppl 1988;24:S37-43. [Pubmed]
    [7]
    Lemke HD, Heidland A, Schaefer RM. Hypersensitivity reactions during haemodialysis: role of complement fragments and ethylene oxide antibodies. Nephrol Dial Transplant 1990;5:264-9. [Pubmed]
    [8]
    Pegues DA, Beck-Sague CM, Woollen SW, Greenspan B, Burns SM, Bland LA, et al. Anaphylactoid reactions associated with reuse of hollow-fiber hemodialyzers and ACE inhibitors. Kidney Int 1992;42:1232-7. [Pubmed]
    [9]
    Centers for Disease Control (CDC). Acute allergic reactions associated with reprocessed hemodialyzers¿United States, 1989-1990. MMWR Morb Mortal Wkly Rep 1991;40(9):147, 153-4.
    [10]
    Tielemans C, Madhoun P, Lenaers M, Schandene L, Goldman M, Vanherweghem JL. Anaphylactoid reactions during hemodialysis on AN69 membranes in patients receiving ACE inhibitors. Kidney Int 1990;38:982-4. [Pubmed]
    [11]
    Alvarez-Lara MA, Martín-Malo A, Espinosa M, Castillo D, Aljama P. ACE inhibitors and anaphilactoid reactions during AN-69 membrane dialysis. Lancet 1991;337(8737):370-1. [Pubmed]
    [12]
    Schulman G, Hakim R, Arias R, Silverberg M, Kaplan AP, Arbeit L. Bradykinin generation by dialysis membranes: possible role in anaphylactic reaction. J Am Soc Nephrol 1993;3:1563-9. [Pubmed]
    [13]
    Verresen L, Fink E, Lemke HD, Vanrenterghem Y. Bradykinin is a mediator of anaphylactoid reactions during hemodialysis with AN69 membranes. Kidney Int 1994;45:1497-503. [Pubmed]
    [14]
    Maheut H, Lacour F. Using AN69 ST membrane: a dialysis centre experience. Nephrol Dial Transplant 2001;16:1519-20. [Pubmed]
    [15]
    Maurice F, Rivory JP, Larsson PH, Johansson SG, Bousquet J. Anaphylactic shock caused by formaldehyde in a patient undergoing long-term hemodialysis. J Allergy Clin Immunol 1986;77:594-7. [Pubmed]
    [16]
    Kalpaklioglu AF, Aydin G. Prevalence of latex sensitivity among patients with chronic renal failure: a new risk group? Artif Organs 1999;23:139-42. [Pubmed]
    [17]
    Ebo DG, Haine SE, Hagendorens MM, Bridts CH, Conraads VM, Vorlat A, et al. Hypersensitivity to nadroparin calcium: case report and review of the literature. Clin Drug Investig 2004;24:421-6. [Pubmed]
    [18]
    Berkun Y, Aviv YS, Schwartz LB, Shalit M. Heparin-induced recurrent anaphylaxis. Clin Exp Allergy 2004;34(12):1916-8. [Pubmed]
    [19]
    Michael B, Coyne DW, Fishbane S, Folkert V, Lynn R, Nissenson AR, et al. Sodium ferric gluconate complex in hemodialysis patients: adverse reactions compared to placebo and iron dextran. Kidney Int 2002;61:1830-9. [Pubmed]
    [20]
    Craddock PR, Fehr J, Dalmasso AP, Brighan KL, Jacob HS. Hemodialysis leukopenia. Pulmonary vascular leukostasis resulting from complement activation by dialyzer cellophane membranes. J Clin Invest 1977;59:879-88. [Pubmed]
    [21]
    Hakim RM, Breillatt J, Lazarus JM, Port FK. Complement activation and hypersensitivity reactions to dialysis membranes. N Engl J Med 1984;311:878-82. [Pubmed]
    [22]
    Aljama P, Martín-Malo A, Castillo D, Velasco F, Torres A, Pérez R, et al. Anaphylatoxin C5a generation and dialysis-induced leukopenia with different hemodialyzer membranes. Blood Purif 1986;4:88-92. [Pubmed]
    [23]
    Martin-Malo A, Castillo D, Castro M, Pérez R, Rios A, Jaraba M, et al. Biocompatibility of dialysis membranes: a comparative study. Nephrol Dial Transplant 1991;6 Suppl 2:55-8.
    [24]
    Canaud B, Aljama P, Tielemans C, Gasparovic V, Gutierrez A, Locatelli F. Pathochemical toxicity of perfluorocarbon-5070, a liquid test performance fluid previously used in dialyzer manufacturing, confirmed in animal experiment. J Am Soc Nephrol 2005;16:1819-2.
    [25]
    Sánchez-Villanueva RJ, González E, Quirce S, Díaz R, Álvarez L, Menéndez D, et al. Reacciones de hipersensibilidad a membranas sintéticas de hemodiálisis. Nefrologia 2014;34(4):520-5. [Pubmed]
    [26]
    Martín-Navarro JA, Gutiérrez-Sánchez MJ, Petkov-Stoyanov V. Hipersensibilidad a membranas sintéticas de hemodiálisis. Nefrologia 2014;34(6):XXX-XX.
    [27]
    Bigazzi R, Atti M, Baldari G. High-permeable membranes and hypersensitivity-like reactions: role of dialysis fluid contamination. Blood Purif 1990;8(4):190-8. [Pubmed]
    [28]
    Muller TF, Seitz M, Eckle I, Lange H, Kolb G. Biocompatibility differences with respect to the dialyzer sterilization method. Nephron 1998;78:139-42. [Pubmed]
    [29]
    Kuragano T, Kuno T, Takahashi Y, Yamamoto C, Nagura Y, Takahashi S, et al. Comparison of the effects of cellulose triacetate and polysulfone membrane on GPIIb/IIIa and platelet activation. Blood Purif 2003;21:176-82. [Pubmed]
    [30]
    Urbani A, Sirolli V, Lupisella S, Levi-Mortera S, Pavone B, Pieroni L, et al. Proteomic investigations on the effect of different membrane materials on blood protein adsorption during haemodialysis. Blood Transfus 2012;10 Suppl 2:s101-12. [Pubmed]
    [31]
    Mares J, Thongboonkerd V, Tuma Z, Moravec J, Matejovic M. Specific adsorption of some complement activation proteins to polysulfone dialysis membranes during hemodialysis. Kidney Int 2009;76:404-13. [Pubmed]
    [32]
    Mares J, Richtrova P, Hricinova A, Tuma Z, Moravec J, Lysak D, et al. Proteomic profiling of blood-dialyzer interactome reveals involvement of lectin complement pathway in hemodialysis-induced inflammatory response. Proteomics Clin Appl 2010;4:829-38. [Pubmed]
    [33]
    Marques ID, Pinheiro KF, de Freitas do Carmo LP, Costa MC, Abensur H. Anaphylactic reaction induced by a polysulfone/polyvinylpyrrolidone membrane in the 10th session of hemodialysis with the same dialyzer. Hemodial Int 2011;15(3):399-403. [Pubmed]
    [34]
    Arenas MD, Gil MT, Carreton MA, Moledous A, Albiach B. Eventos adversos a polisulfona durante la sesión de hemodiálisis. Nefrologia 2007;27:638-42. [Pubmed]
    [35]
    Huang WH, Lee YY, Shih LC. Delayed near-fatal anaphylactic reaction induced by the F10-HPS polysulphone haemodialyser. Nephrol Dial Transplant 2008;23(1):423-4. [Pubmed]
    [36]
    Arenas MD, Niveiro E, Moledous A, Gil MT, Albiach B, Carretón MA. Fatal acute systemic hypersensitivity reaction during haemodialysis. Nephrol Dial Transplant 2006;21(10):2966-70. [Pubmed]
    [37]
    Heegard KD, Tilley MA, Stewart IJ, Edgecombe HP, Lundy JB, Renz EM, et al. Anaphylactoid reaction during first hemofiltration with a PUREMA polysulfone membrane. Int J Artif Organs 2013;36(5):363-6. [Pubmed]
    [38]
    Ohashi N, Yonemura K, Goto T, Suzuki H, Fujigaki Y, Yamamoto T, et al. A case of anaphylactoid shock induced by the BS polysulfone hemodialyzer but not by the F8-HPS polysulfone hemodialyzer. Clin Nephrol 2003;60:214-7. [Pubmed]
    [39]
    Tanoue N, Nagano K, Matsumura H. Use of a light-polymerized composite removable partial denture base for a patient hypersensitive to poly(methyl methacrylate), polysulfone, and polycarbonate: a clinical report. J Prosthet Dent 2005;93:17-20. [Pubmed]
    Descargar PDF
    Idiomas
    Nefrología
    Opciones de artículo
    Herramientas
    es en

    ¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

    Are you a health professional able to prescribe or dispense drugs?