array:20 [
  "pii" => "X0211699505019457"
  "issn" => "02116995"
  "estado" => "S300"
  "fechaPublicacion" => "2005-12-01"
  "documento" => "article"
  "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
  "subdocumento" => "fla"
  "cita" => "Nefrologia. 2005;25:678-83"
  "abierto" => array:3 [
    "ES" => true
    "ES2" => true
    "LATM" => true
  ]
  "gratuito" => true
  "lecturas" => array:2 [
    "total" => 6299
    "formatos" => array:3 [
      "EPUB" => 302
      "HTML" => 5002
      "PDF" => 995
    ]
  ]
  "Traduccion" => array:1 [
    "en" => array:16 [
      "pii" => "X2013251405019454"
      "issn" => "20132514"
      "estado" => "S300"
      "fechaPublicacion" => "2005-12-01"
      "documento" => "article"
      "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
      "subdocumento" => "fla"
      "cita" => "Nefrologia (English Version). 2005;25:678-83"
      "abierto" => array:3 [
        "ES" => true
        "ES2" => true
        "LATM" => true
      ]
      "gratuito" => true
      "lecturas" => array:2 [
        "total" => 2273
        "formatos" => array:3 [
          "EPUB" => 152
          "HTML" => 1895
          "PDF" => 226
        ]
      ]
      "en" => array:7 [
        "idiomaDefecto" => true
        "titulo" => "Study of vascular access (VA) by color Doppler ultrasonography (CDU). Comparison between delta-H and CDU methods in measuring VA blood flow rate"
        "tieneTextoCompleto" => 0
        "tieneResumen" => "en"
        "paginas" => array:1 [
          0 => array:2 [
            "paginaInicial" => "678"
            "paginaFinal" => "683"
          ]
        ]
        "contieneResumen" => array:1 [
          "en" => true
        ]
        "autores" => array:1 [
          0 => array:2 [
            "autoresLista" => "R. Roca-Tey, A. Rivas, R. Samon, O. Ibrik, J. Viladoms"
            "autores" => array:1 [
              0 => array:1 [
                "nombre" => "R. Roca-Tey, A. Rivas, R. Samon, O. Ibrik, J. Viladoms"
              ]
            ]
          ]
        ]
      ]
      "idiomaDefecto" => "en"
      "Traduccion" => array:1 [
        "es" => array:8 [
          "pii" => "X0211699505019457"
          "estado" => "S300"
          "subdocumento" => ""
          "abierto" => array:3 [
            "ES" => true
            "ES2" => true
            "LATM" => true
          ]
          "gratuito" => true
          "lecturas" => array:1 [
            "total" => 0
          ]
          "idiomaDefecto" => "es"
          "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699505019457?idApp=UINPBA000064"
        ]
      ]
      "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X2013251405019454?idApp=UINPBA000064"
      "url" => "/20132514/0000002500000006/v0_201502091553/X2013251405019454/v0_201502091553/en/main.assets"
    ]
  ]
  "itemSiguiente" => array:16 [
    "pii" => "X0211699505019359"
    "issn" => "02116995"
    "estado" => "S300"
    "fechaPublicacion" => "2005-12-01"
    "documento" => "article"
    "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
    "subdocumento" => "fla"
    "cita" => "Nefrologia. 2005;25:684-97"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 15181
      "formatos" => array:3 [
        "EPUB" => 319
        "HTML" => 9359
        "PDF" => 5503
      ]
    ]
    "es" => array:9 [
      "idiomaDefecto" => true
      "titulo" => "ESTUDIO MULTIVARIANTE DE LOS FACTORES PSICOSOCIALES QUE INFLUYEN EN LA ACTITUD POBLACIONAL HACIA LA DONACIÓN DE ÓRGANOS"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "684"
          "paginaFinal" => "697"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Multivariate study of the psychosocial factors affecting public attitude towards organ donation"
        ]
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "C. Conesa, A. Ríos, P. Ramírez, M. Canteras, M.M. Rodríguez, P. Parrilla"
          "autores" => array:1 [
            0 => array:1 [
              "nombre" => "C. Conesa, A. Ríos, P. Ramírez, M. Canteras, M.M. Rodríguez, P. Parrilla"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "Traduccion" => array:1 [
      "en" => array:8 [
        "pii" => "X2013251405019356"
        "estado" => "S300"
        "subdocumento" => ""
        "abierto" => array:3 [
          "ES" => true
          "ES2" => true
          "LATM" => true
        ]
        "gratuito" => true
        "lecturas" => array:1 [
          "total" => 0
        ]
        "idiomaDefecto" => "en"
        "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X2013251405019356?idApp=UINPBA000064"
      ]
    ]
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699505019359?idApp=UINPBA000064"
    "url" => "/02116995/0000002500000006/v0_201502091332/X0211699505019359/v0_201502091333/es/main.assets"
  ]
  "itemAnterior" => array:16 [
    "pii" => "X0211699505019473"
    "issn" => "02116995"
    "estado" => "S300"
    "fechaPublicacion" => "2005-12-01"
    "documento" => "article"
    "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
    "subdocumento" => "fla"
    "cita" => "Nefrologia. 2005;25:668-77"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 8674
      "formatos" => array:3 [
        "EPUB" => 295
        "HTML" => 6979
        "PDF" => 1400
      ]
    ]
    "es" => array:9 [
      "idiomaDefecto" => true
      "titulo" => "Insuficiencia cardíaca y disfunción diastólica en pacientes en hemodiálisis. factores asociados"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "668"
          "paginaFinal" => "677"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Cardiac failure and diastolic disfunction in hemodialysis patients: associated factors"
        ]
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "M. Salgueira, J.A. Milán, R. Moreno Alba, J. Amor, N. Aresté, E. Jiménez, A. Palma"
          "autores" => array:1 [
            0 => array:1 [
              "nombre" => "M. Salgueira, J.A. Milán, R. Moreno Alba, J. Amor, N. Aresté, E. Jiménez, A. Palma"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "Traduccion" => array:1 [
      "en" => array:8 [
        "pii" => "X2013251405019470"
        "estado" => "S300"
        "subdocumento" => ""
        "abierto" => array:3 [
          "ES" => true
          "ES2" => true
          "LATM" => true
        ]
        "gratuito" => true
        "lecturas" => array:1 [
          "total" => 0
        ]
        "idiomaDefecto" => "en"
        "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X2013251405019470?idApp=UINPBA000064"
      ]
    ]
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699505019473?idApp=UINPBA000064"
    "url" => "/02116995/0000002500000006/v0_201502091332/X0211699505019473/v0_201502091332/es/main.assets"
  ]
  "es" => array:10 [
    "idiomaDefecto" => true
    "titulo" => "Estudio del acceso vascular (AV) mediante ecografía Doppler color (EDC). Comparación entre los métod"
    "tieneTextoCompleto" => true
    "paginas" => array:1 [
      0 => array:2 [
        "paginaInicial" => "678"
        "paginaFinal" => "683"
      ]
    ]
    "autores" => array:1 [
      0 => array:2 [
        "autoresLista" => "R. Roca-Tey, A. Rivas, R. Samon, O. Ibrik, J. Viladoms"
        "autores" => array:1 [
          0 => array:1 [
            "nombre" => "R. Roca-Tey, A. Rivas, R. Samon, O. Ibrik, J. Viladoms"
          ]
        ]
      ]
    ]
    "titulosAlternativos" => array:1 [
      "en" => array:1 [
        "titulo" => "Study of vascular access (VA) by color Doppler ultrasonography (CDU). Comparison between delta-H and CDU methods in measuring VA blood flow rate"
      ]
    ]
    "textoCompleto" => "NEFROLOGÍA. Volumen 25. Número 6. 2005 Estudio del acceso vascular (AV) mediante ecografía Doppler color (EDC). Comparación entre los métodos EDC y delta-H aplicados para la determinación del flujo sanguíneo del AV R. Roca-Tey, A. Rivas*, R. Samon, O. Ibrik y J. Viladoms Servicio de Nefrología del Hospital de Mollet. *CETIR Centre Mèdic. Barcelona. RESUMEN Introducción: La ecografía Doppler color (EDC) es una técnica no invasiva de monitorización del acceso vascular (AV) que proporciona información tanto anatómica como funcional del mismo. Por otra parte, la aplicación del método de monitorización delta-H ( H) permite determinar de forma indirecta el flujo sanguíneo del acceso vascular durante la hemodiálisis (HD). Objetivos: 1) Analizar algunos parámetros morfológicos y funcionales del AV mediante EDC. 2) Efectuar un estudio comparativo entre los métodos EDC y H en la determinación del flujo sanguíneo del AV. Material y métodos: Hemos explorado durante 3 meses el AV (fístula arteriovenosa FAVI radial 60,6%, FAVI humeral 24,2%, prótesis femoral 15,2%&#59; duración media 33,5 ± 31,5 meses desde su construción) mediante EDC en 33 pacientes (pts) (edad media 61,7 ± 13,3 años&#59; sexo H: 48,5%&#59; M: 51,5%&#59; tiempo en HD: 39,3 ± 33,3 meses&#59; 9,1% diabetes) afectos de IRC. Todas las exploraciones se efectuaron fuera de la sesión de HD por el mismo radiólogo utilizando un transductor lineal de 5-10 MHz. El flujo sanguíneo arterial y venoso del AV se calculó por EDC utilizando la fórmula: flujo (ml/min) = curva tiempo-velocidad (media de 3 ciclos cardíacos) (m/s) × área transversal (mm2) × 60. Antes de iniciar la evaluación ecográfica, ningún AV presentaba signos de disfunción. Todos los pts con estenosis significativa del AV por EDC (reducción 50% del diámetro endoluminal) se remitieron para efectuar angiografía (AG). Todos los parámetros funcionales se determinaron por duplicado (los 2 valores obtenidos se promediaron). Además, el flujo sanguíneo del AV (QA) se determinó durante la misma semana mediante el método H (monitor Crit-Line III). Resultados: 1) Arteria nutricia (AN). Diámetro medio (dAN): 0,7 ± 0,2 cm. Velocidad máxima sistólica: 161,8 ± 44,5 cm/s. Flujo sanguíneo medio (flujo AN) (n = 27): 2.030,8 ± 987,1 ml/min. Hemos objetivado una correlación (R2) positiva entre flujoAN y dAN (R2 = 0,39, p < 0,001). 2) Vena arterializada (VA). Diámetro medio (dVA): 0,7 ± 0,2 cm. Flujo sanguíneo medio (flujo VA): 1.783,8 ± 1009,7 ml/min (intervalo, 398-5.843 ml/min). Coeficiente de variación medio para medidas duplicadas de flujo VA: 9,2%. Hemos hallado una correlación positiva entre flujo VA y las siguientes variables: dAN (R2 = 0,22, p = 0,005), flujo AN (R2 = 0,19, p = 0,022) y dVA (R2 = 0,14, p = 0,034). Se evidenció estenosis significativa del AV por ECD en 3 pts (9,1%) y en todos ellos la topografía y el grado de estenosis se confirmaron por AG. El flujo VA fue significativamente inferior en los AV con este- Correspondencia: Dr. Ramón Roca Tey Miembro de la SEN Viladomat, 49 08015 Barcelona E-mail: 18647rrt@comb.es 678 MONITORIZACIÓN DEL ACCESO VASCULAR MEDIANTE ECOGRAFÍA DOPPLER COLOR nosis (511,0 ± 179,1 ml/min) respecto a los AV sin estenosis (1.911,1 ± 968,7 ml/min) (p = 0,006). Mediante el análisis de concordancia diagnóstica, hemos objectivado una alta correlación entre los valores de flujo sanguíneo del AV obtenidos por los métodos EDC (flujo VA) y H (QA medio), considerando todas las determinaciones de flujo VA (n = 33, CCI = 0,74, p < 0,0001), para flujo VA < 2.000 ml/min (n = 23, CCI = 0,82, p < 0,0001) y para flujo VA < 1.500 ml/min (n = 14, CCI = 0,73, p = 0,001). No hemos hallado diferencias al comparar QA medio (1593,8 ± 645,7 ml/min, intervalo 559-2.778 ml/min) y flujo VA (p = 0,082). Conclusiones: 1) La EDC es una técnica eficaz y reproducible para la determinación del flujo sanguíneo del AV y para el diagnóstico de la estenosis subclínica del mismo. 2) El flujo sanguíneo del AV está en relación con los parámetros obtenidos en la arteria nutricia. 3) Los métodos EDC y delta-H son superponibles cuando se aplican en la determinación del flujo sanguíneo del AV. Palabras clave: Ecografía Doppler color. Monitorización del acceso vascular. Método delta-H. Crit-Line III. Estenosis del acceso vascular. STUDY OF VASCULAR ACCESS (VA) BY COLOR DOPPLER ULTRASONOGRAPHY (CDU). COMPARISON BETWEEN DELTA-H AND CDU METHODS IN MEASURING VA BLOOD FLOW RATE SUMMARY Introduction: The color Doppler ultrasonography (CDU) is a noninvasive vascular access (VA) monitoring technique that provides both structural and hemodynamic information from VA. On the other hand, the delta-H ( H) method is another noninvasive technique that measures the VA blood flow rate during hemodialysis (HD). Objective: 1) To analyze some anatomic and functional parameters of VA by CDU. 2) To compare H and CDU methods in measuring VA blood flow rate. Patients and methods: We explored the VA (radial arteriovenous fistula AVF 60.6%, brachial AVF 24.2%, femoral graft 15.2%&#59; mean VA duration 33.5 ± 31.5 months) of 33 stable ESRD (mean age 61.7 ± 13.3 yr, sex M: 48.5%&#59; F: 51.5%&#59; mean time on hemodialysis HD 39.3 ± 33.3 months&#59; 9.1% diabetes) patients (pts) by CDU over 3 months period. All CDU examinations were performed off HD by the same radiologist using a 5-10 MHz linear transducer. The arterial and venous blood flow rates were calculated by CDU using the equation: flow (ml/min) = time average velocity (mean of three cardiac cycles) (m/s) × cross-sectional area (mm2) × 60. No VA explored had any sign of dysfunction prior to CDU evaluation. All pts with significant VA stenosis by CDU ( 50% reduction in the luminal diameter) were referred for angiography (AG). All functional parameters were measured in duplicate and the values were averaged. In addition, QA was also calculated in the same week by the H method during HD using the Crit Line III Monitor. Results: Feeding artery (FA) results. Mean FA diameter (FAd): 0.7 ± 0.2 cm. Mean peak systolic velocity: 161.8 ± 44.5 m/s. Mean FA blood flow rate (FA flow) (n = 27): 2,030.8 ± 987.1 ml/min. We found a positive correlation between mean FA flow and mean FAd (R2 = 0.39, p < 0.001). Arterialized vein (AV) results. Mean AV diameter (AVd): 0.7 ± 0.2 cm. Mean AV blood flow rate (AVflow): 1,783.8 ± 1,009.7 ml/min (range, 398-5,843 ml/min). Mean coefficient of variation for duplicate AV flow measurements: 9.2%. We found a positive correlation between mean AV flow and: mean FAd (R2 = 0.22, p = 0.005), mean FAflow (R2 = 0.19, p = 0.022) and mean AVd (R2 = 0.14, p = 0.034). Three pts (9.1%) showed significant VA stenosis by CDU and in, all cases, the location and degree of stenoses were confirmed by AG. The mean AVflow was lower in VA with significant stenosis (511.0 ± 179.1 ml/min) compared to VA without stenosis (1,911.1 ± 968.7 ml/min) (p = 0.006). The calcula679 R. ROCA-TEY y cols. ted values of VA blood flow rate obtained by the CDU technique (AV flow) were highly correlated with those determined by the H method (QA) when considering all comparison values (n = 33, ICC = 0.74, p < 0.0001), for AV flow < 2,000 ml/min (n = 23, ICC = 0.82, p < 0.0001) and for AV flow < 1,500 ml/min (n = 14, ICC = 0.73, p = 0.001). No significant difference was found when QA (1,593.8 ± 645.7 ml/min, range 559-2,778 ml/min) and AV flow were compared (p = 0.082). Conclusions: 1) The CDU technique is a valuable and reproducible method for AV flow measurement and for early diagnosis of significant VA stenosis. 2) The AV flow is related to FA parameters. 3) Both methods, CDU and H, correlated highly with each other when were applied on VA blood flow rate measurement. Key words: Color Doppler ultrasonography. Vascular access surveillance. DeltaH method. Crit-Line III. Vascular access stenosis. INTRODUCCIÓN La ecografía Doppler color (EDC) es una técnica de monitorización del acceso vascular (AV) de segunda generación que presenta algunas ventajas adicionales sobre otros métodos de seguimiento1. Además de proporcionarnos información hemodinámica del AV (determinación directa del flujo sanguíneo), es una técnica de imagen no invasiva que nos permite visualizar cambios morfológicos en el mismo (detección y cuantificación de la estenosis)2, 3. La EDC también se ha utilizado con éxito como soporte de imagen para el tratamiento electivo mediante angioplastia de la estenosis del AV4. Por otra parte, en los últimos años, se han introducido diversos métodos de screening que permiten la monitorización funcional el AV mediante la determinación indirecta del flujo sanguíneo (QA) del mismo5. Uno de ellos, es el método delta-H o H (también denominado técnica por dilución del hematocrito o de ultrafiltración), que hemos implantado en nuestro Servicio desde el año 2000 y que posibilita determinar durante la hemodiálisis (HD) el QA del AV de forma periódica6. En el presente trabajo, se ha efectuado un estudio observacional, transversal y comparativo mediante EDC con los siguientes objetivos: 1) Analizar mediante EDC diversos parámetros morfológicos y funcionales del AV. 2) Efectuar una comparación funcional entre las técnicas CDU y H en la determinación del flujo sanguíneo del AV. MATERIAL Y MÉTODOS Pacientes El AV se evaluó durante 3 meses (agosto-octubre del 2003) en 33 pacientes afectos de insuficiencia 680 renal en programa de HD crónica. Todos los enfermos efectuaban HD tres veces por semana en el Servicio de Nefrología del Hospital de Mollet a través de un AV permanente y canalizado por bipunción. Ningún paciente presentaba su AV con signos de disfunción antes de iniciar el estudio. La tabla I resume las características clínicas más importantes de los enfermos explorados. Métodos Exploración del AV mediante EDC. El AV se evaluó en los planos longitudinal y transversal mediante un transductor lineal de 5-10 MHz utilizando el monitor de ECD Aspen (Siemens-Acuson, Mountain View, CA). Ángulo de exploración: 60º. Todas las exploraciones ecográficas se efectuaron fuera de la sesión de HD por el mismo radiólogo. El flujo sanguíneo arterial y venoso del AV se calculó por EDC utilizando la fórmula: flujo (ml/min) = curva tiempovelocidad (media de 3 ciclos cardíacos) (m/s) × área transversal (mm2) × 60. Los pacientes con estenosis significativa del AV (reducción 50% del diámetro endoluminal) por EDC se remitieron para efectuar angiografía (AG)&#59; no se efectuó AG al resto de casos. Todos los parámetros funcionales se determinaron por duplicado y los 2 valores obtenidos se promediaron. Determinación del flujo sanguíneo del AV mediante el método delta-H. El flujo sanguíneo del AV (QA) se determinó durante la misma semana mediante el método H (monitor Crit-Line III). Este método, descrito y validado por Yarar y cols.7, es una técnica fotométrica que se fundamenta en la relación inversa existente entre la volemia y el hematocrito arterial (Hto). El QA se determina durante la primera hora de la sesión de HD a partir de los cambios del Hto en relación a cambios bruscos de la UF MONITORIZACIÓN DEL ACCESO VASCULAR MEDIANTE ECOGRAFÍA DOPPLER COLOR Tabla I. Características de los pacientes explorados mediante EDC Sexo (%) Edad media (años) Tiempo medio en HD (meses) Diabetes mellitus (%) Tipo de AV (%) Duración media del AV (meses) AV previos Ratio número AV/paciente Hombre 48,5 / Mujer 51,5 61,7 ± 13,3 (intervalo, 31-87) 39,3 ± 33,3 (intervalo, 2-136) 9,1 FAVI radial 60,6 o humeral 24,2&#59; prótesis femoral 15,2 33,5 ± 31,5 (intervalo, 5-113) 39,4% (13/33) pacientes 1,8 ± 1,4 (intervalo, 1-7) RESULTADOS 1) Exploración de la arteria nutricia (AN). El diámetro medio (dAN), la velocidad máxima sistólica y el flujo sanguíneo medio (flujo AN, n = 27) de la AN fueron 0,7 ± 0,2 cm, 161,8 ± 44,5 cm/s y 2.030,8 ± 987,1 ml/min, respectivamente. Hemos objetivado una correlación positiva entre flujo AN y dAN (R2 = 0,39, p < 0,001). 2) Exploración de la vena arterializada (VA). El diámetro medio (dVA) y el flujo sanguíneo medio (flujo VA) de la VA fueron 0,7 ± 0,2 cm y 1.783,8 ± 1.009,7 ml/min (intervalo, 398-5.843 ml/min), respectivamente. Se ha registrado una buena correlación entre las medidas duplicadas de flujo VA con una coeficiente de variación medio para las mismas de 9,2% (0,092 ± 0,10). Hemos hallado una correlación positiva entre flujo VA y las siguientes variables: dAN (R2 = 0,22, p = 0,005), flujo AN (R2 = 0,19, p = 0,022) y dVA (R2 = 0,14, p = 0,034). A pesar que el flujo VA fue inferior en los pacientes ancianos ( 65 años, 1.446,9 ± 729,2 ml/min) respecto a los jóvenes (edad < 65 años, 2.064,5 ± 1.139,1 ml/min) así como en los enfermos diabéticos (1.000,5 ± 573,8 ml/min) respecto a los no-diabéticos (1.862,1 ± 1.016,1 ml/min), las diferencias no han sido estadísticamente significativas (p = 0,06 y p = 0,09, respectivamente)&#59; ninguna diferencia en relación a flujo VA al comparar hombres con mujeres (p = 0,75) o FAVI con prótesis (p = 0,84). Tres enfermos (9,1%) presentaron una estenosis significativa del AV por ECD localizadas en: 1) anastomosis venosa de una prótesis&#59; 2) en la VA de una FAVI radial a 2,5 cm de la anastomosis, y 3) en ambas ramas arterial y venosa alrededor de la anastomosis de otra FAVI radial&#59; el grado medio de estenosis fue del 62,5%. En los 3 casos, la topografía y el grado de estenosis se confirmaron mediante AG. El flujo VA fue significativamente inferior en los AV con estenosis diagnosticada por ECD (511,0 ± 179,1 ml/min) respecto a los AV sin estenosis (1.911,1 ± 968,7 ml/min) (p = 0,006). Mediante la aplicación del análisis de concordancia diagnóstica, hemos objectivado una alta correlación entre los valores de flujo sanguíneo del AV obtenidos por los métodos EDC (flujo VA) y H (QA medio), considerando todas las determinaciones de flujo VA (n = 33, CCI = 0,74, p < 0,0001, fig. I), para flujo VA < 2.000 ml/min (n = 23, CCI = 0,82, p < 0,0001) y para flujo VA < 1.500 ml/min (n = 14, CCI = 0,73, p = 0,001). No hemos hallado diferencias al comparar QA medio (1.593,8 ± 645,7 ml/min, intervalo 559-2.778 ml/min) y flujo VA (p = 0,082). 681 (desde 0,1 hasta 1,8 L/h) con las líneas de HD en configuración normal e invertida. Los cambios de Hto son registrados continuamente por un sensor óptico que se acopla a una cámara sanguínea insertada entre el dializador y la línea arterial. El QA se calculó mediante la siguiente fórmula: QA = (UF máx-UF mín) · Hto máx inv/ Hto inv- Hto nor donde UF máx es la ultrafiltración máxima, UF mín es la ultrafiltración mínima, Hto máx inv es el Hto máximo obtenido con las líneas de HD en posición invertida, Hto inv es el cambio en el Hto arterial con las líneas invertidas, y Hto nor es el cambio en el Hto arterial con las líneas de HD en situación normal. Estudio estadístico El análisis estadístico de los datos se efectuó con el programa SPSS versión 9.0 para Windows. Los valores se expresaron como porcentajes o media ± desviación estándar. Las diferencias entre medias independientes se analizaron utilizando el test de Student para datos no pareados y el análisis de varianza (ANOVA). Para evaluar las diferencias entre medias relacionadas se utilizó el test de Student para datos pareados. Para el análisis de correlación entre el flujo sanguíneo medio del AV y las distintas variables cuantitativas estudiadas se calculó el coeficiente de correlación de Pearson. Para evaluar las diferencias entre variables cualitativas se utilizó el test de chicuadrado. Se ha efectuado el análisis de concordancia diagnóstica mediante el cálculo del coeficiente de correlación intraclase (CCI) entre los valores de flujo sanguíneo del AV aplicando los métodos EDC y H (factor fijo) sobre la misma muestra de individuos (factor aleatorio). Se ha considerado estadísticamente significativo una p < 0,05. R. ROCA-TEY y cols. Gráfico global (n = 33) 6.000 5.000 4.000 3.000 2.000 1.000 0 0 1.000 Delta-H Fig. 1.--Análisis de concordancia diagnóstica. Se ha calculado el coeficiente de correlación intraclase entre los valores de flujo sanguíneo del AV obtenidos mediante los métodos Doppler-color y Delta-H (n = 33, CCI = 0,74, p < 0,0001). 2.000 3.000 DISCUSIÓN En el presente estudio, la exploración del AV mediante EDC nos ha permitido diagnosticar 3 casos de estenosis significativa del AV no sospechados previamente y que se confirmaron todos ellos mediante AG. Otros autores también han objetivado la eficacia de la EDC como método de screening en el diagnóstico precoz de la estenosis anatómica tanto de FAVI o como de prótesis de PTFE8, 9. Además, se ha demostrado una correlación lineal entre la EDC como técnica de imagen no invasiva y la AG como método invasivo de diagnóstico respecto al grado de estenosis subclínica del AV8. En el estudio de May y cols., efectuado en pacientes dializados a través de prótesis de PTFE, los AV con un grado de estenosis visualizada por EDC superior al 70% presentaron la máxima prevalencia de trombosis durante el periodo de estudio10. En nuestro trabajo, la EDC se ha mostrado como una técnica válida y reproducible en la determinación del flujo sanguíneo del AV (flujo VA). El coeficiente de variación para medidas duplicadas de flujo VA obtenido por EDC ha sido inferior al 10% y es comparable al obtenido tras aplicar otros métodos de estimación del flujo sanguíneo del AV7, 11-13. En nuestro estudio, los AV con estenosis significativa objetivada mediante EDC y, por tanto, con 682 mayor riesgo de trombosis, presentaban un flujo VA significativamente inferior respecto a los AV sin estenosis. En este sentido, algunos trabajos han constatado que los AV explorados por EDC que se trombosaron o fueron sometidos a intervención correctiva durante el periodo de seguimiento presentaban inicialmente un flujo VA significativamente inferior respecto al resto de AV10, 14, 15. En la serie de Miranda y Sands, sólo el 26% de las FAVI exploradas con EDC con flujo VA 400 ml/min estaban permeables y sin intervención al finalizar los 6 meses de seguimiento16. Por su parte, en el estudio de Bay y cols., el riesgo relativo de trombosis o intervención de prótesis de PTFE monitorizadas por EDC fue del 40% en caso de flujo VA inicial inferior a 500 ml/min y este riesgo se duplicó cuando flujo VA fue inferior a 300 ml/min17. En el presente trabajo, el flujo VA se ha correlacionado con los parámetros obtenidos al explorar la arteria nutricia (diámetro y flujo sanguíneo arteriales). En la serie de Wiese y cols., referido a 15 pts dializados a través de prótesis de PTFE braquial, el flujo sanguíneo de la arteria nutricia obtenido por EDC en el periodo postoperatorio inmediato fue significativamente inferior en los AV que se trombosaron durante los siguientes 15 meses respecto al resto de AV18. En nuestro estudio, se ha puesto de manifiesto la concordancia funcional entre los métodos EDC y Delta-H. No hemos hallado diferencias significativas al comparar los valores de flujo sanguíneo del AV obtenidos por ambos métodos. Otros autores han presentado resultados similares tras comparar los valores de flujo sanguíneo del AV obtenidos mediante la EDC y el método indirecto de dilución ultrasónica9, 19, 20. En resumen, los resultados del presente estudio confirman las prestaciones de la EDC, tanto desde el punto de vista anatómico como funcional, y justifican su inclusión en las estrategias actuales de monitorización del AV. BIBLIOGRAFÍA 1. Hakim R & Himmelfarb J: Hemodialysis access failure: a call to action. Kidney Int 54: 1029-1040, 1998. 2. Lockhart ME & Robbin ML: Hemodialysis access ultrasound. Ultrasound Quarterly 17: 157-167, 2001. 3. Sands JJ, Ferrell LM, Perry MA: The role of Color Flow Doppler Ultrasound in Dialysis Access. Seminars in Nephrology 22: 195-201, 2002. 4. Bacchini G, Cappello A, La Milia V y cols.: Color Doppler ultrasonography imaging to guide transluminal angioplasty of venous stenosis. Kidney Int 58: 1810-1813, 2000. 5. Lindsay RM & Leypoldt JK: Monitoring vascular access flow. Advances in Renal Replacement Therapy 6: 273-277, 1999. 6. Roca-Tey R, Samon R, Ibrik O y cols.: Monitorización del acceso vascular mediante la determinación del flujo sanguíneo Doppler color MONITORIZACIÓN DEL ACCESO VASCULAR MEDIANTE ECOGRAFÍA DOPPLER COLOR 7. 8. 9. 10. 11. 12. 13. durante la hemodiálisis por el método de ultrafiltración. Estudio prospectivo de 65 pacientes. Nefrologia XXIV: 246-252, 2004. Yarar D, Cheung AK, Sakiewicz P y cols.: Ultrafiltration method for measuring vascular access flow rates during hemodialysis. Kidney Int 56: 1129-1135, 1999. Gadallah MF, Paulson WD, Vickers B, Work J: Accuracy of Doppler Ultrasound in diagnosing anatomic stenosis of hemodialysis arteriovenous access as compared with fistulography. Am J Kidney Dis 32: 273-277, 1998. Older RA, Gizienski TA, Wilkowski MJ, Angle JF, Cote DA: Hemodialysis access stenosis: early detection with color Doppler US. Radiology 207: 161-164, 1998. May RE, Himmelfarb J, Yenicesu M y cols.: Predictive measures of vascular access thrombosis: a prospective study. Kidney Int 52: 1656-1662, 1997. Pandeya S & Lindsay RM: The relationship between cardiac output and access flow during hemodialysis. ASAIO Journal 45: 135-138, 1999. Roca-Tey R, Samon R, Ibrik O, Viladoms J: Estudio del flujo sanguíneo (QA) del acceso vascular (AV) durante la hemodiálisis (HD) en 64 pacientes mediante el método de ultrafiltración. Nefrologia XXI (Supl. 2): 44. Abstract, 2001. Lindsay RM, Blake PG, Malek P y cols.: Hemodialysis access blood flow rates can be measured by a Differential Conduc- 14. 15. 16. 17. 18. 19. 20. tivity technique and are predictive of access clotting. Am J Kidney Dis 30: 475-482, 1997. Shackleton CR, Taylor DC, Buckley AR y cols.: Predicting failure in polytetrafluoroethylene vascular access grafts for hemodialysis: a pilot study. Can J Surg 30: 442-444, 1987. Ok Kim Y, Woo Yang C, Ae Yoon S y cols.: Access blood flow as a predictor of early failures of native arteriovenous fistulas in hemodialysis patients. Am J Nephrol 21: 221-225, 2001. Miranda CL & Sands JJ: Flow volumes as a predictor of hemodialysis access failure. J Vasc Technol 21 155-159, 1997. Bay WH, Henry ML, Lazarus JM y cols.: Predicting hemodialysis access failure with color flow Doppler ultrasound. Am J Nephrol 18: 296-304, 1998. Wiese P, Blume J, Mueller H, Renner H, Nonnast-Daniel B: Clinical and Doppler ultrasonography data of a polyurethane vascular access graft for hemodilaysis: a prospective study. Nephrol Dial Transplant 18: 1397-1400, 2003. Sands J, Glidden D & Miranda C: Hemodialysis access flow measurement. Comparison of ultrasound dilution and Duplex ultrasonography. ASAIO Journal 42: M899-M901, 1996. Schwarz C, Mitterbauer C, Boczula M y cols.: Flow monitoring: performance characteristics of ultrasound dilution versus color Doppler ultrasound compared with with fistulography. Am J Kidney Dis 42: 539-545, 2003. 683 "
    "pdfFichero" => "P1-E249-S132-A1164.pdf"
    "tienePdf" => true
    "PalabrasClave" => array:1 [
      "es" => array:1 [
        0 => array:4 [
          "clase" => "keyword"
          "titulo" => "Palabras clave"
          "identificador" => "xpalclavsec431137"
          "palabras" => array:4 [
            0 => "Ecografia Doppler-color"
            1 => "Monitorización estenosis"
            2 => "Método delta-H"
            3 => "Crit-Line III"
          ]
        ]
      ]
    ]
  ]
  "idiomaDefecto" => "es"
  "url" => "/02116995/0000002500000006/v0_201502091332/X0211699505019457/v0_201502091332/es/main.assets"
  "Apartado" => array:4 [
    "identificador" => "35393"
    "tipo" => "SECCION"
    "es" => array:2 [
      "titulo" => "Artículos Originales"
      "idiomaDefecto" => true
    ]
    "idiomaDefecto" => "es"
  ]
  "PDF" => "https://static.elsevier.es/multimedia/02116995/0000002500000006/v0_201502091332/X0211699505019457/v0_201502091332/es/P1-E249-S132-A1164.pdf?idApp=UINPBA000064&text.app=https://revistanefrologia.com/"
  "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699505019457?idApp=UINPBA000064"
]
Compartir
Información de la revista

Estadísticas

Siga este enlace para acceder al texto completo del artículo

Estudio del acceso vascular (AV) mediante ecografía Doppler color (EDC). Comparación entre los métod
Study of vascular access (VA) by color Doppler ultrasonography (CDU). Comparison between delta-H and CDU methods in measuring VA blood flow rate
R. Roca-Tey, A. Rivas, R. Samon, O. Ibrik, J. Viladoms
Leído
13274
Veces
se ha leído el artículo
3169
Total PDF
10105
Total HTML
Compartir estadísticas
 array:20 [
  "pii" => "X0211699505019457"
  "issn" => "02116995"
  "estado" => "S300"
  "fechaPublicacion" => "2005-12-01"
  "documento" => "article"
  "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
  "subdocumento" => "fla"
  "cita" => "Nefrologia. 2005;25:678-83"
  "abierto" => array:3 [
    "ES" => true
    "ES2" => true
    "LATM" => true
  ]
  "gratuito" => true
  "lecturas" => array:2 [
    "total" => 6299
    "formatos" => array:3 [
      "EPUB" => 302
      "HTML" => 5002
      "PDF" => 995
    ]
  ]
  "Traduccion" => array:1 [
    "en" => array:16 [
      "pii" => "X2013251405019454"
      "issn" => "20132514"
      "estado" => "S300"
      "fechaPublicacion" => "2005-12-01"
      "documento" => "article"
      "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
      "subdocumento" => "fla"
      "cita" => "Nefrologia (English Version). 2005;25:678-83"
      "abierto" => array:3 [
        "ES" => true
        "ES2" => true
        "LATM" => true
      ]
      "gratuito" => true
      "lecturas" => array:2 [
        "total" => 2273
        "formatos" => array:3 [
          "EPUB" => 152
          "HTML" => 1895
          "PDF" => 226
        ]
      ]
      "en" => array:7 [
        "idiomaDefecto" => true
        "titulo" => "Study of vascular access (VA) by color Doppler ultrasonography (CDU). Comparison between delta-H and CDU methods in measuring VA blood flow rate"
        "tieneTextoCompleto" => 0
        "tieneResumen" => "en"
        "paginas" => array:1 [
          0 => array:2 [
            "paginaInicial" => "678"
            "paginaFinal" => "683"
          ]
        ]
        "contieneResumen" => array:1 [
          "en" => true
        ]
        "autores" => array:1 [
          0 => array:2 [
            "autoresLista" => "R. Roca-Tey, A. Rivas, R. Samon, O. Ibrik, J. Viladoms"
            "autores" => array:1 [
              0 => array:1 [
                "nombre" => "R. Roca-Tey, A. Rivas, R. Samon, O. Ibrik, J. Viladoms"
              ]
            ]
          ]
        ]
      ]
      "idiomaDefecto" => "en"
      "Traduccion" => array:1 [
        "es" => array:8 [
          "pii" => "X0211699505019457"
          "estado" => "S300"
          "subdocumento" => ""
          "abierto" => array:3 [
            "ES" => true
            "ES2" => true
            "LATM" => true
          ]
          "gratuito" => true
          "lecturas" => array:1 [
            "total" => 0
          ]
          "idiomaDefecto" => "es"
          "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699505019457?idApp=UINPBA000064"
        ]
      ]
      "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X2013251405019454?idApp=UINPBA000064"
      "url" => "/20132514/0000002500000006/v0_201502091553/X2013251405019454/v0_201502091553/en/main.assets"
    ]
  ]
  "itemSiguiente" => array:16 [
    "pii" => "X0211699505019359"
    "issn" => "02116995"
    "estado" => "S300"
    "fechaPublicacion" => "2005-12-01"
    "documento" => "article"
    "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
    "subdocumento" => "fla"
    "cita" => "Nefrologia. 2005;25:684-97"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 15181
      "formatos" => array:3 [
        "EPUB" => 319
        "HTML" => 9359
        "PDF" => 5503
      ]
    ]
    "es" => array:9 [
      "idiomaDefecto" => true
      "titulo" => "ESTUDIO MULTIVARIANTE DE LOS FACTORES PSICOSOCIALES QUE INFLUYEN EN LA ACTITUD POBLACIONAL HACIA LA DONACIÓN DE ÓRGANOS"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "684"
          "paginaFinal" => "697"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Multivariate study of the psychosocial factors affecting public attitude towards organ donation"
        ]
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "C. Conesa, A. Ríos, P. Ramírez, M. Canteras, M.M. Rodríguez, P. Parrilla"
          "autores" => array:1 [
            0 => array:1 [
              "nombre" => "C. Conesa, A. Ríos, P. Ramírez, M. Canteras, M.M. Rodríguez, P. Parrilla"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "Traduccion" => array:1 [
      "en" => array:8 [
        "pii" => "X2013251405019356"
        "estado" => "S300"
        "subdocumento" => ""
        "abierto" => array:3 [
          "ES" => true
          "ES2" => true
          "LATM" => true
        ]
        "gratuito" => true
        "lecturas" => array:1 [
          "total" => 0
        ]
        "idiomaDefecto" => "en"
        "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X2013251405019356?idApp=UINPBA000064"
      ]
    ]
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699505019359?idApp=UINPBA000064"
    "url" => "/02116995/0000002500000006/v0_201502091332/X0211699505019359/v0_201502091333/es/main.assets"
  ]
  "itemAnterior" => array:16 [
    "pii" => "X0211699505019473"
    "issn" => "02116995"
    "estado" => "S300"
    "fechaPublicacion" => "2005-12-01"
    "documento" => "article"
    "licencia" => "http://www.elsevier.com/open-access/userlicense/1.0/"
    "subdocumento" => "fla"
    "cita" => "Nefrologia. 2005;25:668-77"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 8674
      "formatos" => array:3 [
        "EPUB" => 295
        "HTML" => 6979
        "PDF" => 1400
      ]
    ]
    "es" => array:9 [
      "idiomaDefecto" => true
      "titulo" => "Insuficiencia cardíaca y disfunción diastólica en pacientes en hemodiálisis. factores asociados"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "668"
          "paginaFinal" => "677"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Cardiac failure and diastolic disfunction in hemodialysis patients: associated factors"
        ]
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "M. Salgueira, J.A. Milán, R. Moreno Alba, J. Amor, N. Aresté, E. Jiménez, A. Palma"
          "autores" => array:1 [
            0 => array:1 [
              "nombre" => "M. Salgueira, J.A. Milán, R. Moreno Alba, J. Amor, N. Aresté, E. Jiménez, A. Palma"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "Traduccion" => array:1 [
      "en" => array:8 [
        "pii" => "X2013251405019470"
        "estado" => "S300"
        "subdocumento" => ""
        "abierto" => array:3 [
          "ES" => true
          "ES2" => true
          "LATM" => true
        ]
        "gratuito" => true
        "lecturas" => array:1 [
          "total" => 0
        ]
        "idiomaDefecto" => "en"
        "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X2013251405019470?idApp=UINPBA000064"
      ]
    ]
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699505019473?idApp=UINPBA000064"
    "url" => "/02116995/0000002500000006/v0_201502091332/X0211699505019473/v0_201502091332/es/main.assets"
  ]
  "es" => array:10 [
    "idiomaDefecto" => true
    "titulo" => "Estudio del acceso vascular (AV) mediante ecografía Doppler color (EDC). Comparación entre los métod"
    "tieneTextoCompleto" => true
    "paginas" => array:1 [
      0 => array:2 [
        "paginaInicial" => "678"
        "paginaFinal" => "683"
      ]
    ]
    "autores" => array:1 [
      0 => array:2 [
        "autoresLista" => "R. Roca-Tey, A. Rivas, R. Samon, O. Ibrik, J. Viladoms"
        "autores" => array:1 [
          0 => array:1 [
            "nombre" => "R. Roca-Tey, A. Rivas, R. Samon, O. Ibrik, J. Viladoms"
          ]
        ]
      ]
    ]
    "titulosAlternativos" => array:1 [
      "en" => array:1 [
        "titulo" => "Study of vascular access (VA) by color Doppler ultrasonography (CDU). Comparison between delta-H and CDU methods in measuring VA blood flow rate"
      ]
    ]
    "textoCompleto" => "NEFROLOGÍA. Volumen 25. Número 6. 2005 Estudio del acceso vascular (AV) mediante ecografía Doppler color (EDC). Comparación entre los métodos EDC y delta-H aplicados para la determinación del flujo sanguíneo del AV R. Roca-Tey, A. Rivas*, R. Samon, O. Ibrik y J. Viladoms Servicio de Nefrología del Hospital de Mollet. *CETIR Centre Mèdic. Barcelona. RESUMEN Introducción: La ecografía Doppler color (EDC) es una técnica no invasiva de monitorización del acceso vascular (AV) que proporciona información tanto anatómica como funcional del mismo. Por otra parte, la aplicación del método de monitorización delta-H ( H) permite determinar de forma indirecta el flujo sanguíneo del acceso vascular durante la hemodiálisis (HD). Objetivos: 1) Analizar algunos parámetros morfológicos y funcionales del AV mediante EDC. 2) Efectuar un estudio comparativo entre los métodos EDC y H en la determinación del flujo sanguíneo del AV. Material y métodos: Hemos explorado durante 3 meses el AV (fístula arteriovenosa FAVI radial 60,6%, FAVI humeral 24,2%, prótesis femoral 15,2%&#59; duración media 33,5 ± 31,5 meses desde su construción) mediante EDC en 33 pacientes (pts) (edad media 61,7 ± 13,3 años&#59; sexo H: 48,5%&#59; M: 51,5%&#59; tiempo en HD: 39,3 ± 33,3 meses&#59; 9,1% diabetes) afectos de IRC. Todas las exploraciones se efectuaron fuera de la sesión de HD por el mismo radiólogo utilizando un transductor lineal de 5-10 MHz. El flujo sanguíneo arterial y venoso del AV se calculó por EDC utilizando la fórmula: flujo (ml/min) = curva tiempo-velocidad (media de 3 ciclos cardíacos) (m/s) × área transversal (mm2) × 60. Antes de iniciar la evaluación ecográfica, ningún AV presentaba signos de disfunción. Todos los pts con estenosis significativa del AV por EDC (reducción 50% del diámetro endoluminal) se remitieron para efectuar angiografía (AG). Todos los parámetros funcionales se determinaron por duplicado (los 2 valores obtenidos se promediaron). Además, el flujo sanguíneo del AV (QA) se determinó durante la misma semana mediante el método H (monitor Crit-Line III). Resultados: 1) Arteria nutricia (AN). Diámetro medio (dAN): 0,7 ± 0,2 cm. Velocidad máxima sistólica: 161,8 ± 44,5 cm/s. Flujo sanguíneo medio (flujo AN) (n = 27): 2.030,8 ± 987,1 ml/min. Hemos objetivado una correlación (R2) positiva entre flujoAN y dAN (R2 = 0,39, p < 0,001). 2) Vena arterializada (VA). Diámetro medio (dVA): 0,7 ± 0,2 cm. Flujo sanguíneo medio (flujo VA): 1.783,8 ± 1009,7 ml/min (intervalo, 398-5.843 ml/min). Coeficiente de variación medio para medidas duplicadas de flujo VA: 9,2%. Hemos hallado una correlación positiva entre flujo VA y las siguientes variables: dAN (R2 = 0,22, p = 0,005), flujo AN (R2 = 0,19, p = 0,022) y dVA (R2 = 0,14, p = 0,034). Se evidenció estenosis significativa del AV por ECD en 3 pts (9,1%) y en todos ellos la topografía y el grado de estenosis se confirmaron por AG. El flujo VA fue significativamente inferior en los AV con este- Correspondencia: Dr. Ramón Roca Tey Miembro de la SEN Viladomat, 49 08015 Barcelona E-mail: 18647rrt@comb.es 678 MONITORIZACIÓN DEL ACCESO VASCULAR MEDIANTE ECOGRAFÍA DOPPLER COLOR nosis (511,0 ± 179,1 ml/min) respecto a los AV sin estenosis (1.911,1 ± 968,7 ml/min) (p = 0,006). Mediante el análisis de concordancia diagnóstica, hemos objectivado una alta correlación entre los valores de flujo sanguíneo del AV obtenidos por los métodos EDC (flujo VA) y H (QA medio), considerando todas las determinaciones de flujo VA (n = 33, CCI = 0,74, p < 0,0001), para flujo VA < 2.000 ml/min (n = 23, CCI = 0,82, p < 0,0001) y para flujo VA < 1.500 ml/min (n = 14, CCI = 0,73, p = 0,001). No hemos hallado diferencias al comparar QA medio (1593,8 ± 645,7 ml/min, intervalo 559-2.778 ml/min) y flujo VA (p = 0,082). Conclusiones: 1) La EDC es una técnica eficaz y reproducible para la determinación del flujo sanguíneo del AV y para el diagnóstico de la estenosis subclínica del mismo. 2) El flujo sanguíneo del AV está en relación con los parámetros obtenidos en la arteria nutricia. 3) Los métodos EDC y delta-H son superponibles cuando se aplican en la determinación del flujo sanguíneo del AV. Palabras clave: Ecografía Doppler color. Monitorización del acceso vascular. Método delta-H. Crit-Line III. Estenosis del acceso vascular. STUDY OF VASCULAR ACCESS (VA) BY COLOR DOPPLER ULTRASONOGRAPHY (CDU). COMPARISON BETWEEN DELTA-H AND CDU METHODS IN MEASURING VA BLOOD FLOW RATE SUMMARY Introduction: The color Doppler ultrasonography (CDU) is a noninvasive vascular access (VA) monitoring technique that provides both structural and hemodynamic information from VA. On the other hand, the delta-H ( H) method is another noninvasive technique that measures the VA blood flow rate during hemodialysis (HD). Objective: 1) To analyze some anatomic and functional parameters of VA by CDU. 2) To compare H and CDU methods in measuring VA blood flow rate. Patients and methods: We explored the VA (radial arteriovenous fistula AVF 60.6%, brachial AVF 24.2%, femoral graft 15.2%&#59; mean VA duration 33.5 ± 31.5 months) of 33 stable ESRD (mean age 61.7 ± 13.3 yr, sex M: 48.5%&#59; F: 51.5%&#59; mean time on hemodialysis HD 39.3 ± 33.3 months&#59; 9.1% diabetes) patients (pts) by CDU over 3 months period. All CDU examinations were performed off HD by the same radiologist using a 5-10 MHz linear transducer. The arterial and venous blood flow rates were calculated by CDU using the equation: flow (ml/min) = time average velocity (mean of three cardiac cycles) (m/s) × cross-sectional area (mm2) × 60. No VA explored had any sign of dysfunction prior to CDU evaluation. All pts with significant VA stenosis by CDU ( 50% reduction in the luminal diameter) were referred for angiography (AG). All functional parameters were measured in duplicate and the values were averaged. In addition, QA was also calculated in the same week by the H method during HD using the Crit Line III Monitor. Results: Feeding artery (FA) results. Mean FA diameter (FAd): 0.7 ± 0.2 cm. Mean peak systolic velocity: 161.8 ± 44.5 m/s. Mean FA blood flow rate (FA flow) (n = 27): 2,030.8 ± 987.1 ml/min. We found a positive correlation between mean FA flow and mean FAd (R2 = 0.39, p < 0.001). Arterialized vein (AV) results. Mean AV diameter (AVd): 0.7 ± 0.2 cm. Mean AV blood flow rate (AVflow): 1,783.8 ± 1,009.7 ml/min (range, 398-5,843 ml/min). Mean coefficient of variation for duplicate AV flow measurements: 9.2%. We found a positive correlation between mean AV flow and: mean FAd (R2 = 0.22, p = 0.005), mean FAflow (R2 = 0.19, p = 0.022) and mean AVd (R2 = 0.14, p = 0.034). Three pts (9.1%) showed significant VA stenosis by CDU and in, all cases, the location and degree of stenoses were confirmed by AG. The mean AVflow was lower in VA with significant stenosis (511.0 ± 179.1 ml/min) compared to VA without stenosis (1,911.1 ± 968.7 ml/min) (p = 0.006). The calcula679 R. ROCA-TEY y cols. ted values of VA blood flow rate obtained by the CDU technique (AV flow) were highly correlated with those determined by the H method (QA) when considering all comparison values (n = 33, ICC = 0.74, p < 0.0001), for AV flow < 2,000 ml/min (n = 23, ICC = 0.82, p < 0.0001) and for AV flow < 1,500 ml/min (n = 14, ICC = 0.73, p = 0.001). No significant difference was found when QA (1,593.8 ± 645.7 ml/min, range 559-2,778 ml/min) and AV flow were compared (p = 0.082). Conclusions: 1) The CDU technique is a valuable and reproducible method for AV flow measurement and for early diagnosis of significant VA stenosis. 2) The AV flow is related to FA parameters. 3) Both methods, CDU and H, correlated highly with each other when were applied on VA blood flow rate measurement. Key words: Color Doppler ultrasonography. Vascular access surveillance. DeltaH method. Crit-Line III. Vascular access stenosis. INTRODUCCIÓN La ecografía Doppler color (EDC) es una técnica de monitorización del acceso vascular (AV) de segunda generación que presenta algunas ventajas adicionales sobre otros métodos de seguimiento1. Además de proporcionarnos información hemodinámica del AV (determinación directa del flujo sanguíneo), es una técnica de imagen no invasiva que nos permite visualizar cambios morfológicos en el mismo (detección y cuantificación de la estenosis)2, 3. La EDC también se ha utilizado con éxito como soporte de imagen para el tratamiento electivo mediante angioplastia de la estenosis del AV4. Por otra parte, en los últimos años, se han introducido diversos métodos de screening que permiten la monitorización funcional el AV mediante la determinación indirecta del flujo sanguíneo (QA) del mismo5. Uno de ellos, es el método delta-H o H (también denominado técnica por dilución del hematocrito o de ultrafiltración), que hemos implantado en nuestro Servicio desde el año 2000 y que posibilita determinar durante la hemodiálisis (HD) el QA del AV de forma periódica6. En el presente trabajo, se ha efectuado un estudio observacional, transversal y comparativo mediante EDC con los siguientes objetivos: 1) Analizar mediante EDC diversos parámetros morfológicos y funcionales del AV. 2) Efectuar una comparación funcional entre las técnicas CDU y H en la determinación del flujo sanguíneo del AV. MATERIAL Y MÉTODOS Pacientes El AV se evaluó durante 3 meses (agosto-octubre del 2003) en 33 pacientes afectos de insuficiencia 680 renal en programa de HD crónica. Todos los enfermos efectuaban HD tres veces por semana en el Servicio de Nefrología del Hospital de Mollet a través de un AV permanente y canalizado por bipunción. Ningún paciente presentaba su AV con signos de disfunción antes de iniciar el estudio. La tabla I resume las características clínicas más importantes de los enfermos explorados. Métodos Exploración del AV mediante EDC. El AV se evaluó en los planos longitudinal y transversal mediante un transductor lineal de 5-10 MHz utilizando el monitor de ECD Aspen (Siemens-Acuson, Mountain View, CA). Ángulo de exploración: 60º. Todas las exploraciones ecográficas se efectuaron fuera de la sesión de HD por el mismo radiólogo. El flujo sanguíneo arterial y venoso del AV se calculó por EDC utilizando la fórmula: flujo (ml/min) = curva tiempovelocidad (media de 3 ciclos cardíacos) (m/s) × área transversal (mm2) × 60. Los pacientes con estenosis significativa del AV (reducción 50% del diámetro endoluminal) por EDC se remitieron para efectuar angiografía (AG)&#59; no se efectuó AG al resto de casos. Todos los parámetros funcionales se determinaron por duplicado y los 2 valores obtenidos se promediaron. Determinación del flujo sanguíneo del AV mediante el método delta-H. El flujo sanguíneo del AV (QA) se determinó durante la misma semana mediante el método H (monitor Crit-Line III). Este método, descrito y validado por Yarar y cols.7, es una técnica fotométrica que se fundamenta en la relación inversa existente entre la volemia y el hematocrito arterial (Hto). El QA se determina durante la primera hora de la sesión de HD a partir de los cambios del Hto en relación a cambios bruscos de la UF MONITORIZACIÓN DEL ACCESO VASCULAR MEDIANTE ECOGRAFÍA DOPPLER COLOR Tabla I. Características de los pacientes explorados mediante EDC Sexo (%) Edad media (años) Tiempo medio en HD (meses) Diabetes mellitus (%) Tipo de AV (%) Duración media del AV (meses) AV previos Ratio número AV/paciente Hombre 48,5 / Mujer 51,5 61,7 ± 13,3 (intervalo, 31-87) 39,3 ± 33,3 (intervalo, 2-136) 9,1 FAVI radial 60,6 o humeral 24,2&#59; prótesis femoral 15,2 33,5 ± 31,5 (intervalo, 5-113) 39,4% (13/33) pacientes 1,8 ± 1,4 (intervalo, 1-7) RESULTADOS 1) Exploración de la arteria nutricia (AN). El diámetro medio (dAN), la velocidad máxima sistólica y el flujo sanguíneo medio (flujo AN, n = 27) de la AN fueron 0,7 ± 0,2 cm, 161,8 ± 44,5 cm/s y 2.030,8 ± 987,1 ml/min, respectivamente. Hemos objetivado una correlación positiva entre flujo AN y dAN (R2 = 0,39, p < 0,001). 2) Exploración de la vena arterializada (VA). El diámetro medio (dVA) y el flujo sanguíneo medio (flujo VA) de la VA fueron 0,7 ± 0,2 cm y 1.783,8 ± 1.009,7 ml/min (intervalo, 398-5.843 ml/min), respectivamente. Se ha registrado una buena correlación entre las medidas duplicadas de flujo VA con una coeficiente de variación medio para las mismas de 9,2% (0,092 ± 0,10). Hemos hallado una correlación positiva entre flujo VA y las siguientes variables: dAN (R2 = 0,22, p = 0,005), flujo AN (R2 = 0,19, p = 0,022) y dVA (R2 = 0,14, p = 0,034). A pesar que el flujo VA fue inferior en los pacientes ancianos ( 65 años, 1.446,9 ± 729,2 ml/min) respecto a los jóvenes (edad < 65 años, 2.064,5 ± 1.139,1 ml/min) así como en los enfermos diabéticos (1.000,5 ± 573,8 ml/min) respecto a los no-diabéticos (1.862,1 ± 1.016,1 ml/min), las diferencias no han sido estadísticamente significativas (p = 0,06 y p = 0,09, respectivamente)&#59; ninguna diferencia en relación a flujo VA al comparar hombres con mujeres (p = 0,75) o FAVI con prótesis (p = 0,84). Tres enfermos (9,1%) presentaron una estenosis significativa del AV por ECD localizadas en: 1) anastomosis venosa de una prótesis&#59; 2) en la VA de una FAVI radial a 2,5 cm de la anastomosis, y 3) en ambas ramas arterial y venosa alrededor de la anastomosis de otra FAVI radial&#59; el grado medio de estenosis fue del 62,5%. En los 3 casos, la topografía y el grado de estenosis se confirmaron mediante AG. El flujo VA fue significativamente inferior en los AV con estenosis diagnosticada por ECD (511,0 ± 179,1 ml/min) respecto a los AV sin estenosis (1.911,1 ± 968,7 ml/min) (p = 0,006). Mediante la aplicación del análisis de concordancia diagnóstica, hemos objectivado una alta correlación entre los valores de flujo sanguíneo del AV obtenidos por los métodos EDC (flujo VA) y H (QA medio), considerando todas las determinaciones de flujo VA (n = 33, CCI = 0,74, p < 0,0001, fig. I), para flujo VA < 2.000 ml/min (n = 23, CCI = 0,82, p < 0,0001) y para flujo VA < 1.500 ml/min (n = 14, CCI = 0,73, p = 0,001). No hemos hallado diferencias al comparar QA medio (1.593,8 ± 645,7 ml/min, intervalo 559-2.778 ml/min) y flujo VA (p = 0,082). 681 (desde 0,1 hasta 1,8 L/h) con las líneas de HD en configuración normal e invertida. Los cambios de Hto son registrados continuamente por un sensor óptico que se acopla a una cámara sanguínea insertada entre el dializador y la línea arterial. El QA se calculó mediante la siguiente fórmula: QA = (UF máx-UF mín) · Hto máx inv/ Hto inv- Hto nor donde UF máx es la ultrafiltración máxima, UF mín es la ultrafiltración mínima, Hto máx inv es el Hto máximo obtenido con las líneas de HD en posición invertida, Hto inv es el cambio en el Hto arterial con las líneas invertidas, y Hto nor es el cambio en el Hto arterial con las líneas de HD en situación normal. Estudio estadístico El análisis estadístico de los datos se efectuó con el programa SPSS versión 9.0 para Windows. Los valores se expresaron como porcentajes o media ± desviación estándar. Las diferencias entre medias independientes se analizaron utilizando el test de Student para datos no pareados y el análisis de varianza (ANOVA). Para evaluar las diferencias entre medias relacionadas se utilizó el test de Student para datos pareados. Para el análisis de correlación entre el flujo sanguíneo medio del AV y las distintas variables cuantitativas estudiadas se calculó el coeficiente de correlación de Pearson. Para evaluar las diferencias entre variables cualitativas se utilizó el test de chicuadrado. Se ha efectuado el análisis de concordancia diagnóstica mediante el cálculo del coeficiente de correlación intraclase (CCI) entre los valores de flujo sanguíneo del AV aplicando los métodos EDC y H (factor fijo) sobre la misma muestra de individuos (factor aleatorio). Se ha considerado estadísticamente significativo una p < 0,05. R. ROCA-TEY y cols. Gráfico global (n = 33) 6.000 5.000 4.000 3.000 2.000 1.000 0 0 1.000 Delta-H Fig. 1.--Análisis de concordancia diagnóstica. Se ha calculado el coeficiente de correlación intraclase entre los valores de flujo sanguíneo del AV obtenidos mediante los métodos Doppler-color y Delta-H (n = 33, CCI = 0,74, p < 0,0001). 2.000 3.000 DISCUSIÓN En el presente estudio, la exploración del AV mediante EDC nos ha permitido diagnosticar 3 casos de estenosis significativa del AV no sospechados previamente y que se confirmaron todos ellos mediante AG. Otros autores también han objetivado la eficacia de la EDC como método de screening en el diagnóstico precoz de la estenosis anatómica tanto de FAVI o como de prótesis de PTFE8, 9. Además, se ha demostrado una correlación lineal entre la EDC como técnica de imagen no invasiva y la AG como método invasivo de diagnóstico respecto al grado de estenosis subclínica del AV8. En el estudio de May y cols., efectuado en pacientes dializados a través de prótesis de PTFE, los AV con un grado de estenosis visualizada por EDC superior al 70% presentaron la máxima prevalencia de trombosis durante el periodo de estudio10. En nuestro trabajo, la EDC se ha mostrado como una técnica válida y reproducible en la determinación del flujo sanguíneo del AV (flujo VA). El coeficiente de variación para medidas duplicadas de flujo VA obtenido por EDC ha sido inferior al 10% y es comparable al obtenido tras aplicar otros métodos de estimación del flujo sanguíneo del AV7, 11-13. En nuestro estudio, los AV con estenosis significativa objetivada mediante EDC y, por tanto, con 682 mayor riesgo de trombosis, presentaban un flujo VA significativamente inferior respecto a los AV sin estenosis. En este sentido, algunos trabajos han constatado que los AV explorados por EDC que se trombosaron o fueron sometidos a intervención correctiva durante el periodo de seguimiento presentaban inicialmente un flujo VA significativamente inferior respecto al resto de AV10, 14, 15. En la serie de Miranda y Sands, sólo el 26% de las FAVI exploradas con EDC con flujo VA 400 ml/min estaban permeables y sin intervención al finalizar los 6 meses de seguimiento16. Por su parte, en el estudio de Bay y cols., el riesgo relativo de trombosis o intervención de prótesis de PTFE monitorizadas por EDC fue del 40% en caso de flujo VA inicial inferior a 500 ml/min y este riesgo se duplicó cuando flujo VA fue inferior a 300 ml/min17. En el presente trabajo, el flujo VA se ha correlacionado con los parámetros obtenidos al explorar la arteria nutricia (diámetro y flujo sanguíneo arteriales). En la serie de Wiese y cols., referido a 15 pts dializados a través de prótesis de PTFE braquial, el flujo sanguíneo de la arteria nutricia obtenido por EDC en el periodo postoperatorio inmediato fue significativamente inferior en los AV que se trombosaron durante los siguientes 15 meses respecto al resto de AV18. En nuestro estudio, se ha puesto de manifiesto la concordancia funcional entre los métodos EDC y Delta-H. No hemos hallado diferencias significativas al comparar los valores de flujo sanguíneo del AV obtenidos por ambos métodos. Otros autores han presentado resultados similares tras comparar los valores de flujo sanguíneo del AV obtenidos mediante la EDC y el método indirecto de dilución ultrasónica9, 19, 20. En resumen, los resultados del presente estudio confirman las prestaciones de la EDC, tanto desde el punto de vista anatómico como funcional, y justifican su inclusión en las estrategias actuales de monitorización del AV. BIBLIOGRAFÍA 1. Hakim R & Himmelfarb J: Hemodialysis access failure: a call to action. Kidney Int 54: 1029-1040, 1998. 2. Lockhart ME & Robbin ML: Hemodialysis access ultrasound. Ultrasound Quarterly 17: 157-167, 2001. 3. Sands JJ, Ferrell LM, Perry MA: The role of Color Flow Doppler Ultrasound in Dialysis Access. Seminars in Nephrology 22: 195-201, 2002. 4. Bacchini G, Cappello A, La Milia V y cols.: Color Doppler ultrasonography imaging to guide transluminal angioplasty of venous stenosis. Kidney Int 58: 1810-1813, 2000. 5. Lindsay RM & Leypoldt JK: Monitoring vascular access flow. Advances in Renal Replacement Therapy 6: 273-277, 1999. 6. Roca-Tey R, Samon R, Ibrik O y cols.: Monitorización del acceso vascular mediante la determinación del flujo sanguíneo Doppler color MONITORIZACIÓN DEL ACCESO VASCULAR MEDIANTE ECOGRAFÍA DOPPLER COLOR 7. 8. 9. 10. 11. 12. 13. durante la hemodiálisis por el método de ultrafiltración. Estudio prospectivo de 65 pacientes. Nefrologia XXIV: 246-252, 2004. Yarar D, Cheung AK, Sakiewicz P y cols.: Ultrafiltration method for measuring vascular access flow rates during hemodialysis. Kidney Int 56: 1129-1135, 1999. Gadallah MF, Paulson WD, Vickers B, Work J: Accuracy of Doppler Ultrasound in diagnosing anatomic stenosis of hemodialysis arteriovenous access as compared with fistulography. Am J Kidney Dis 32: 273-277, 1998. Older RA, Gizienski TA, Wilkowski MJ, Angle JF, Cote DA: Hemodialysis access stenosis: early detection with color Doppler US. Radiology 207: 161-164, 1998. May RE, Himmelfarb J, Yenicesu M y cols.: Predictive measures of vascular access thrombosis: a prospective study. Kidney Int 52: 1656-1662, 1997. Pandeya S & Lindsay RM: The relationship between cardiac output and access flow during hemodialysis. ASAIO Journal 45: 135-138, 1999. Roca-Tey R, Samon R, Ibrik O, Viladoms J: Estudio del flujo sanguíneo (QA) del acceso vascular (AV) durante la hemodiálisis (HD) en 64 pacientes mediante el método de ultrafiltración. Nefrologia XXI (Supl. 2): 44. Abstract, 2001. Lindsay RM, Blake PG, Malek P y cols.: Hemodialysis access blood flow rates can be measured by a Differential Conduc- 14. 15. 16. 17. 18. 19. 20. tivity technique and are predictive of access clotting. Am J Kidney Dis 30: 475-482, 1997. Shackleton CR, Taylor DC, Buckley AR y cols.: Predicting failure in polytetrafluoroethylene vascular access grafts for hemodialysis: a pilot study. Can J Surg 30: 442-444, 1987. Ok Kim Y, Woo Yang C, Ae Yoon S y cols.: Access blood flow as a predictor of early failures of native arteriovenous fistulas in hemodialysis patients. Am J Nephrol 21: 221-225, 2001. Miranda CL & Sands JJ: Flow volumes as a predictor of hemodialysis access failure. J Vasc Technol 21 155-159, 1997. Bay WH, Henry ML, Lazarus JM y cols.: Predicting hemodialysis access failure with color flow Doppler ultrasound. Am J Nephrol 18: 296-304, 1998. Wiese P, Blume J, Mueller H, Renner H, Nonnast-Daniel B: Clinical and Doppler ultrasonography data of a polyurethane vascular access graft for hemodilaysis: a prospective study. Nephrol Dial Transplant 18: 1397-1400, 2003. Sands J, Glidden D & Miranda C: Hemodialysis access flow measurement. Comparison of ultrasound dilution and Duplex ultrasonography. ASAIO Journal 42: M899-M901, 1996. Schwarz C, Mitterbauer C, Boczula M y cols.: Flow monitoring: performance characteristics of ultrasound dilution versus color Doppler ultrasound compared with with fistulography. Am J Kidney Dis 42: 539-545, 2003. 683 "
    "pdfFichero" => "P1-E249-S132-A1164.pdf"
    "tienePdf" => true
    "PalabrasClave" => array:1 [
      "es" => array:1 [
        0 => array:4 [
          "clase" => "keyword"
          "titulo" => "Palabras clave"
          "identificador" => "xpalclavsec431137"
          "palabras" => array:4 [
            0 => "Ecografia Doppler-color"
            1 => "Monitorización estenosis"
            2 => "Método delta-H"
            3 => "Crit-Line III"
          ]
        ]
      ]
    ]
  ]
  "idiomaDefecto" => "es"
  "url" => "/02116995/0000002500000006/v0_201502091332/X0211699505019457/v0_201502091332/es/main.assets"
  "Apartado" => array:4 [
    "identificador" => "35393"
    "tipo" => "SECCION"
    "es" => array:2 [
      "titulo" => "Artículos Originales"
      "idiomaDefecto" => true
    ]
    "idiomaDefecto" => "es"
  ]
  "PDF" => "https://static.elsevier.es/multimedia/02116995/0000002500000006/v0_201502091332/X0211699505019457/v0_201502091332/es/P1-E249-S132-A1164.pdf?idApp=UINPBA000064&text.app=https://revistanefrologia.com/"
  "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/X0211699505019457?idApp=UINPBA000064"
]
Información del artículo
ISSN: 02116995
Idioma original: Español
Datos actualizados diariamente
año/Mes Html Pdf Total
2024 Noviembre 1 2 3
2024 Octubre 78 72 150
2024 Septiembre 54 34 88
2024 Agosto 79 73 152
2024 Julio 76 56 132
2024 Junio 86 48 134
2024 Mayo 115 52 167
2024 Abril 82 46 128
2024 Marzo 88 28 116
2024 Febrero 83 39 122
2024 Enero 58 39 97
2023 Diciembre 79 26 105
2023 Noviembre 95 36 131
2023 Octubre 126 38 164
2023 Septiembre 146 35 181
2023 Agosto 136 28 164
2023 Julio 148 35 183
2023 Junio 166 20 186
2023 Mayo 147 40 187
2023 Abril 91 18 109
2023 Marzo 141 29 170
2023 Febrero 126 34 160
2023 Enero 114 33 147
2022 Diciembre 82 44 126
2022 Noviembre 123 37 160
2022 Octubre 109 39 148
2022 Septiembre 94 41 135
2022 Agosto 104 47 151
2022 Julio 71 52 123
2022 Junio 64 39 103
2022 Mayo 70 47 117
2022 Abril 78 54 132
2022 Marzo 70 48 118
2022 Febrero 74 42 116
2022 Enero 89 53 142
2021 Diciembre 76 52 128
2021 Noviembre 81 41 122
2021 Octubre 82 57 139
2021 Septiembre 101 55 156
2021 Agosto 85 51 136
2021 Julio 78 43 121
2021 Junio 61 30 91
2021 Mayo 63 66 129
2021 Abril 141 58 199
2021 Marzo 80 51 131
2021 Febrero 85 23 108
2021 Enero 61 35 96
2020 Diciembre 58 16 74
2020 Noviembre 52 10 62
2020 Octubre 59 20 79
2020 Septiembre 53 14 67
2020 Agosto 55 25 80
2020 Julio 55 20 75
2020 Junio 72 23 95
2020 Mayo 76 24 100
2020 Abril 118 25 143
2020 Marzo 113 19 132
2020 Febrero 145 20 165
2020 Enero 158 26 184
2019 Diciembre 98 33 131
2019 Noviembre 134 49 183
2019 Octubre 169 62 231
2019 Septiembre 165 66 231
2019 Agosto 121 26 147
2019 Julio 121 58 179
2019 Junio 138 53 191
2019 Mayo 120 50 170
2019 Abril 124 64 188
2019 Marzo 78 35 113
2019 Febrero 42 18 60
2019 Enero 44 24 68
2018 Diciembre 73 31 104
2018 Noviembre 93 13 106
2018 Octubre 63 13 76
2018 Septiembre 81 14 95
2018 Agosto 71 24 95
2018 Julio 40 16 56
2018 Junio 55 14 69
2018 Mayo 45 14 59
2018 Abril 63 9 72
2018 Marzo 53 9 62
2018 Febrero 33 5 38
2018 Enero 50 9 59
2017 Diciembre 49 8 57
2017 Noviembre 52 20 72
2017 Octubre 48 10 58
2017 Septiembre 33 8 41
2017 Agosto 32 14 46
2017 Julio 35 6 41
2017 Junio 40 10 50
2017 Mayo 62 15 77
2017 Abril 47 29 76
2017 Marzo 36 15 51
2017 Febrero 30 11 41
2017 Enero 30 18 48
2016 Diciembre 52 7 59
2016 Noviembre 53 13 66
2016 Octubre 84 20 104
2016 Septiembre 152 17 169
2016 Agosto 216 17 233
2016 Julio 169 14 183
2016 Junio 109 0 109
2016 Mayo 136 0 136
2016 Abril 98 0 98
2016 Marzo 95 0 95
2016 Febrero 107 0 107
2016 Enero 113 0 113
2015 Diciembre 92 0 92
2015 Noviembre 110 0 110
2015 Octubre 106 0 106
2015 Septiembre 117 0 117
2015 Agosto 108 0 108
2015 Julio 104 0 104
2015 Junio 50 0 50
2015 Mayo 97 0 97
2015 Abril 14 0 14
2015 Febrero 4 0 4
Mostrar todo

Siga este enlace para acceder al texto completo del artículo

Idiomas
Nefrología
es en

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?